19.2.2 Direct commands for the tare
Set tare in working memory
#SETT...
#SETT 25.25;g
Store tare in memory
#STOT...
#STOT 1;25.25;g
#STOT 13;452.42;g
#STOT 53;100;g
#STOT 244;200.42;g
Read a tare
#GETT 13
Read all tares
#GETT ALL
The tare weights are printed out in the same format as they are stored with #STOT..
The data received can be stored in a text file and further processed using Excel.
Clear a tare
#CLRT 13
Clear all tares
#CLRT ALL