Numerical Display
The sample weights are available in numerical form on the next page. The results display format can
be selected during the measurement.
Touch the program step in the graph to display it. The active program step is highlighted in red in the
graph, and the number is boxed in red. The requested weights are then automatically displayed in the
table at the end of the time interval (the last ones measured during the current program step).
The temperature curve default can be changed during the measurement. In order to do this, scroll
back to the prepASH user interface and then select the
Program Editor
.
8.6.3 Program End
A sound is emitted to signal the end of the program, provided this setting has been selected in the
program. The ventilation keeps running until the selected temperature has been reached.
CAUTION
The sample chamber hood opens automatically once the temperature selected in the
program has been reached.
Do not place anything on top of the prepASH.
The parts inside the furnace chamber and the metal covers will become hot. Be very
careful to avoid contact.
Always use the supplied holder for removing the sample plate from the device. If you are
removing individual crucibles, always use the supplied tongs (this will also help you to
avoid spurious measurements).