10 Statistics
■
29
En
gl
is
h
10 Statistics
Functions of the statistics program and storage functions
:
„MODE“
In this function-field you may define whether only the statistics-program, only the storage-program or both
programs simultaneously should be used.
„COUNT“
A number of measurements is laid down, after which automatic storage is to be terminated.
„RECORDING“
„MANUAL“:
The «
STO i
» function key must be pressed for each value to be stored.
„LOADCHANGE“: The balance stores the measured value automatically after a load change.
„TIMEBASE“:
The balance stores every value measured after a defined period (default: 2.0 seconds).
„TIMEBASE“
Definition of the time span for the recording of data in accordance with „RECORDING TIMEBASE” (for ex-
ample, every 2.0 seconds).
Assignment of the function keys:
• SET STATISTIC
STATISTIC
MODE
OFF
STATISTIC
RECORDER
STAT./RECORDER
Statistics program off
Statistics only
Data storage only
Statistics and storage
COUNT
100
Number of values to be stored automatically (1..500).
RECORDING
MANUAL
TIMEBASE
LOADCHANGE
With
«
STO i
»
function key
on a time basis
after every weight change
TIMBASE
2.0
Time base for “storage” in seconds
NOTE
In storing the first value a range of ±50% is determined. Subsequent values must be within this range
otherwise an error message will be issued.
Key
Function
«
STO i
»
Take value, start/stop of automatic recording
«
AUTO i
»
Automatic recording is running
«
WAIT i
»
Value not stabilized yet
«
RES
»
Before a new series of measurements the storage must be reset using «
RES
».
The key
must be held down until the acoustic signal sounds and the measurement counter is
reset.
«
END
»
Permanently store data (only with recording activated)
«
INF
»
Change the display to the info display.
Displayed information:
- „Average value (AVERAGE)"
- „Standard deviation (STD DEV.)"
- „Relative standard deviation (STD DEV.-%)"
- „Maximum (MAX)"
- „Minimum (MIN)"
- „individual values"
Press «
esc
» to exit the info display.