Button Print each value:
After pressing the print button once, immediate printout, even if the scale has not yet
released the weight value as stable and thus valid.
•
Button print stable value:
After pressing the print button, one-time printout as soon as the weight value has been
approved by the scale as stable and valid.
•
Continuously:
After pressing the Print button start / stop a continuous printout of the displayed weight in
the speed of the repetition rate, which is set in the weighing parameters,see 6.1.1 Refresh
rate.
•
Time interval:
After pressing the print button start / stop a continuous printout of the displayed weight in
time with the interval.
•
Load change:
After pressing the Print button Start / Stop printout of the displayed weight, whenever a
new weight is placed on the scale and the weight value is approved as stable and valid.
To do this, a load change mode must be set as an absolute value in grams or in% of the
first load.
The layout can be change by pressing "Edit" what is displayed right from”Layout”.
In the edit mode you will find various items to choose from. Please note the text paragraps are also
items and must therefore be placed in the layout to appear at the correct place in the printout.
Example of a measurement block
Summary of Contents for 350-8182-000
Page 1: ...S W I S S M A D E 390 HF HA HE HM Operating Instructions 350 8182 000...
Page 11: ...390 Your new balance Precisa Gravimetrics AG 10 1 6 Assembling Optional Click Click...
Page 12: ...390 Your new balance Precisa Gravimetrics AG 11 45 Click...
Page 13: ...390 Your new balance Precisa Gravimetrics AG 12...
Page 26: ...390 Discover the Capabilities of the series 390 Precisa Gravimetrics AG 25...