390
System Settings
Precisa Gravimetrics AG
57
6.12
Adjustment
Various factors influence the weighing system, e.g. gravity, air temperature, air pressure, humidity,
etc. In order to achieve and guarantee the most accurate weighing results over time, the weighing
system should therefore be routinely readjusted.
Due to the numerous adjustment modes and their settings, you can set the balance so that it
always meets your requirements.
With the automatic adjustment mode, the greatest interference effects are automatically minimized
or. eliminated or the balance at least warn you that an intervention is necessary.
If the Ambient Observation System AOS is installed, the air temperature, air pressure and humidity
can be included, which results in an optimally monitored and precise system, see 12.3Ambient
Observation System AOS (optional).
Parallel to these adjustment modes, the balance can always be adjusted explicitly in the Adjustment
application, see 2.7 Adjusting the balance.
Common settings:
Protocol:
After the adjustment, a protocol is output via the corresponding interface. If the balance is
integrated in an EDP system, this can lead to problems, which is why the output can be
suppressed with the Protocol setting.
Summary of Contents for 350-8182-000
Page 1: ...S W I S S M A D E 390 HF HA HE HM Operating Instructions 350 8182 000...
Page 11: ...390 Your new balance Precisa Gravimetrics AG 10 1 6 Assembling Optional Click Click...
Page 12: ...390 Your new balance Precisa Gravimetrics AG 11 45 Click...
Page 13: ...390 Your new balance Precisa Gravimetrics AG 12...
Page 26: ...390 Discover the Capabilities of the series 390 Precisa Gravimetrics AG 25...