Press "Start"
⇒
the balance automatically adjusts itself by meanings of the built-in self
calibration system (SCS). When this is done it shows the message "Adjustment done".
As soon as calibration has been completed, the message "Calibration completed" appears.
Move left to the application screen and call up "Weighing.
NOTE
The balance must be adjusted on initial installation and after every
relocation.
If you work in accordance with Good Laboratory Practice (GLP), observe
the prescribed intervals between calibrations (adjustments).
For setting all different adjustment modes,
Swipe left for switching the applications screen,
Call up "Settings",
Scroll down, call up "Adjustment",
Press "Adjustment mode" and
Select the mode you wish.
NOTE
Pressing "?" calls up the contextual Help which provides detailled
information about the point you currently are. For quitting Help, press
"Close".