12
13
EEx / I.S. approval:
5531B:
DEMKO 02ATEX132122 ..............................
II 1 G
EEx ib IIC T6
Applicable in zone ....................................... 1 or 2
5531B1:
KEMA 05ATEX1044 X ................................. -5°C
≤
T
a
≤
+60°C
II 3 GD T 85˚C
-20°C
≤
T
a
≤
+60°C
II 3 G
EEx nL IIC T6
Applicable in zone ....................................... 2 or 22
5531B2:
KEMA 05ATEX1105 X ................................. -5°C
≤
T
a
≤
+60°C
II 2 GD T 85˚C
-20°C
≤
T
a
≤
+60°C
II 2 G
EEx ib IIC T6
Applicable in zone ....................................... 1, 2, 21, or 22
EEx / I.S. data:
Ui ................................................................. : 45 VDC
Ii ................................................................... : 500 mA
Pi ................................................................. : 0.9 W
Ci ................................................................. : 0
μ
F
Li .................................................................. : 0 mH
goST R approval:
VNIIM & VNIIFTRI, Cert. no. ........................ See www.prelectronics.com
observed authority requirements:
Standard:
EMC 2004/108/EC ...................................... EN 61326-1
ATEX 94/9/EC .............................................. EN 50014, EN 50020, EN 60079-15,
EN 61241-0 and draftIEC 61241-11
Special conditions for safe use:
Only plugs and cable entries as delivered with the equipment may be used.
For application in explosive atmospheres caused by air/dust mixtures and
where category 3D or 2D appartus is required, the Loop Powered LCD Indicator
type 5531B1 respectively type 5531B2 shall be installed in such a way that
the risk for mechanical damage is low and shall only be used in environments
where the UV light cannot influence the non-metallic parts.
Year of manufacture can be taken from the first two digits of the serial number.
FIELD ENCLoSURE TYPE 8501:
The field enclosure type 8501 complies with industrial standards for efficient
protection of process equipment in wet areas up to IP65. When 8501 is
delivered with a built-in PReview 5531, the assembled unit is approved
for installation in explosive gas atmospheres as well as in explosive dust
environments. The technical specifications of the field enclosure can be found
in the 8501 data sheet.
ELECTRICAL SPECIFICATIoNS:
Specifications range:
-20°C to +60°C
Common specifications:
Max. voltage drop at 20 mA:
Without backlight ........................................ < 1.5 VDC
Half backlight .............................................. < 6.5 VDC
Full backlight ............................................... < 10.5 VDC
Response time (0...90%) ............................. < 1 s
Calibration temperature............................... 20...28°C
Signal dynamics, input ................................ 15 bit
Temperature coefficient ............................... < ±0.01% of span/°C
Linearity error .............................................. < ±0.1% of span
EMC immunity influence ............................. < ±0.5%
Screw terminal torsion ................................ 0.5 Nm
Humidity ...................................................... < 95% (non-cond.)
Dimensions (HxWxD) ................................... 48 x 96 x 120 mm
Cut-out dimensions (HxW) .......................... 44.5 x 91.5 mm
Protection degree (from front) ..................... IP65
Weight ......................................................... 150 g
Input:
Measurement range .................................... 3.6...23 mA
Display:
Display readout ........................................... ±9999 (4 digits)
Min. display readout (span) ......................... 0 counts
Decimal point .............................................. Programmable
Digit height .................................................. 16 mm
Updating speed ........................................... 500 ms
of span = Of the presently selected range