![PP Systems CIRAS-2 Operator'S Manual Download Page 132](http://html1.mh-extra.com/html/pp-systems/ciras-2/ciras-2_operators-manual_1592667132.webp)
Measuring Soil Respiration with CIRAS-2
CIRAS-2 Operator's Manual Version 2.04
- 130 -
Results
Select from two options - save soil respiration (SR) results as i) g m
-2
hr
-1
or
ii) SR as µmol m
-2
s
-1
and
Evaporation (Evap) results as mmol m
-2
s
-1
.
Data File
Used to create new data files. If no file has been selected, you will be prompted when the first records are
taken. Click on
Data File
to create a new data file and enter a name in the File name box (.csc extension
is assumed and does not need to be typed in). Measurements which follow will be saved to this file.
Select File Mode
Determines how data files are managed. There are 3 options:
Always prompt for new Data File
Every time you start a new measurement, a window will appear asking you to enter a new data file name.
Append to Data File
If selected, all measurements will be appended to the existing data file.
Overwrite Data File
If selected, all measurements will overwrite previously stored data in that particular data file.
Treatment
Identify the treatment file used. See
on page 61 for details.
New Measurement
Only available once records have been saved. Allows the specification of a new data file (as well as the
other options).
Graphing Options
Allows setting up the graph min-max Y-axis scales. Click on the Graphing Options button to set the
Minimum and Maximum Y-axis values for reference CO
2
(Cr), Delta CO
2
(dC) and soil respiration rate
(Sr). CO
2
, Delta CO
2
and soil respiration rate are plotted simultaneously.