background image

En

gl

ish

Guidance and manufacture’s declaration – electromagnetic immunity –

for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of 
Pro7HC should assure that it is used in such an environment.

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment - guidance

Electrostatic  

±6 kV contact 

±6 kV contact 

Floors should be wood, concrete or  

discharge (ESD) 

±8 kV air  

±8 kV air 

ceramic tile. If floor are covered with  

IEC 61000-4-2 

 

 

synthetic material, the relative humidity  

   

 

 

should be at least 30%.

Electrical fast  

±2 kV for power 

±2kV for power 

Mains power quality should be that of  

transient/burst 

supply lines 

supply lines 

a typical commercial or hospital   IEC 

61000-4-4  

 

environment.

Surge 

± 1 kV line(s) t 

±1 kV differential 

Mains power quality should be that of  

IEC 61000-4-5 

o line(s) 

mode 

a typical commercial or hospital    

  

 

 

environment.

Voltage dips, short 

<5% U

T

  

<5% U

T

  

Mains power quality should be that  

interruptions and 

(>95% dip in U

T

(>95% dip in U

T

of a typical commercial or hospital  

voltage variations 

for 0.5 cycle 

for 0.5 cycle 

environment. If the user of the Pro7HC  

on power supply 

 

 

requires continued operation during  

input lines 

40% U

T

 

40% U

T

 

power mains interruptions, it is   IEC 

61000-4-11 

(60% dip in U

T

(60% dip in U

T

recommended that the Pro7HC be 

powered  

 

 

for 5 cycles 

for 5 cycles from an 

uninterruptible power supply or a  

 

 

 

 

battery. 

    

70% U

T

 

70% U

T

 

   

(30% dip in U

T

(30% dip in U

T

)

   

for 25 cycles 

for 25 cycles 

   

<5% U

T

  

<5% U

T

  

 

   

(>95% dip in U

T

(>95% dip in U

T

   

for 5 sec 

for 5 sec

Power frequency  

3A/m 

3A/m 

Power frequency magnetic fields should  

(50Hz) magnetic 

 

 

be at levels characteristic of a typical  

61000-4-8 

 

 

location in a typical commercial or  

   

 

 

hospital environment.

 

NOTE   U

T

 is the a.c. mains voltage prior to application of the test level.

13

Guidance and manufacture’s declaration – electromagnetic immunity –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of 
Pro7HC should assure that it is used in such an environment.

Immunity test 

IEC 60601 test 

Compliance  

Electromagnetic environment 

 

 

level 

level  - 

guidance 

   

   

 

 

Portable and mobile RF communications  

   

 

 

equipment should be used no closer to  

   

 

 

any part of the Pro7HC, including cables, than  

   

 

 

the  recommended separation distance  

   

 

 

calculated from the equation applicable to the  

   

 

 

frequency of the transmitter.

   

 

 

Recommended separation distance

Conducted RF 

3 Vrms 

3 Vrms

IEC 61000-4-6 

150 kHz to 80 MHz

Radiated RF 

3 V/m 

3 Vrms 

 

            80 MHz to 800 MHz

IEC 61000-4-3 

80 MHz to 2.5 GHz 

 

 

            800 MHz to 2.5 GHz

  

 

 Where 

P is the maximum output power rating  

   

 

 

of the transmitter in watts (W) according  

   

 

 

to the transmitter manufacturer and d is the  

   

 

 

recommended separation distance in metres (m).

   

 

 

Field strengths from fixed RF transmitters, as  

   

 

 

determined by an electromagnetic site survey,

a

  

   

 

 

should be less than the compliance level in each  

   

 

 

frequency range.

b

   

 

 

Interference may occur in the vicinity of  

   

 

 

equipment marked with the following symbol: 

 
 
 

NOTE 1   At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people.

a

   Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the Pro7HC is used exceeds 
the applicable RF compliance level above, the  Pro7HC should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro7HC.

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

P

d

167

.

1

=

P

d

167

.

1

=

P

d

167

.

1

=

14

Summary of Contents for pro7HC

Page 1: ...English Instructions For Use pro7HCTM Model ...

Page 2: ...st training to your own level and reduce any burden on your joints tendons and ligaments As with every form of training the correct use of exercises adjusted to your personal abilities will determine the benefits and effects of completing a training session on Power Plate while at the same time reducing the risk of injury to the body We recommend con sulting your physician before beginning any new...

Page 3: ...nes CirculationandCardiovascular Claim Power Plate training can be a beneficial tool intervention to improve and increase circulation and improve the function of the cardiovascular system Definition The circulatory system is an organ system that moves nutrients gases and wastes to and from cells helps fight diseases and helps stabilise body temperature and pH to maintain homeostasis Two types of f...

Page 4: ...lug before attempting to retrieve the object If the object cannot be reached contact an authorised Power Plate service professional UseOfThePowerPlate Machine 1 0Generalhealthwarningsforuse Always follow the directions on the machine s console for proper operation Close supervision is required when the machine is used by or near children or disabled persons Always use care when getting on and off ...

Page 5: ...ications equipment transmitters and the pro7HC model as recommended below according to the maximum output power of the communications equipment The adequate separation distance to provide base immunity to RF disturbances is 1 0m Retainthisusermanualforyourfuture reference 08 09 Classification 1 Class I equipment 2 Type B applied part 3 IPX0 4 Not category AP APG equipment 5 Mode of operation Non c...

Page 6: ...g your training As with any training routine it is important to start using the proMOTION on the lowest level of resistance and with the Power Plate machine on a low setting Establish a good technique before increasing the setting on the proMOTION from low to high Always pay special attention to your posture especially the position of your wrists and back and how close you are to the Power Plate m...

Page 7: ...t LIFE SUPPORTING Guidance and manufacture s declaration electromagnetic immunity The Pro7HC is intended for use in the electromagnetic environment specified below The customer or the user of Pro7HC should assure that it is used in such an environment Immunity test IEC 60601 test Compliance Electromagnetic environment level level guidance Portable and mobile RF communications equipment should be u...

Page 8: ...d or any other circumstances influencing the body and mind In general the following progression steps can be taken Exercisetime numberofexercisesand resttime These settings depend completely on your training goals If you are new to Acceleration Training sessions we recommend performing only a few exercises and resting for the same amount of time as is spent actively using the Power Plate machine A...

Page 9: ...However our most important advice is to always listen to what your own body is telling you Changing your exercises by using more movements of your body and that also correspond to your specific goals will help increase performance even more You can think of increasing the angle of the joint which increases the contraction of the muscle or varying the exercises by performing them with a variety of ...

Page 10: ...makes a rattling noise unplug the machine wait at least one minute then check to make sure that the feet are properly tightened DANGER Power Plate strongly recommends that the user not try to resolve any electrical issues regarding the power source to the machine by oneself Instead power plate strongly recommends that the user contact a licensed professional electrician to conduct any examination ...

Page 11: ...l contact details Warranty Product Registration To register your Power Plate machine please visit us online at www powerplate com Copies of this manual and any other paperwork may be obtained by contacting Power Plate through the Contact Us page online or by writing to Power Plate Performance Health Systems UK Ltd The Perfume Factory G 11 140 Wales Farm Road London W3 6UG 44 20 7317 5000 info powe...

Reviews: