Guidance and
manufacture’s
declaration
–
electromagnetic emissions-
for all ME EQUIPMENT and ME SYSTEMS
Guidance and
manufacture’s
declaration
–
electromagnetic emission
The Pro7 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Pro7 should assure that it is used in such an environment.
Wi-FiSpecifications
Network Standard: IEEE 802.11b/g/n
Band: 2.400-2.4835GHz
Cover Range: 20meter
Wi-Fi Power:10dBm
EMC Declaration
Emission test
Compliance
Electromagnetic environment
–
guidance
Group 1
The Pro7 use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interfe rence in nearby electronic
equipment.
RF emission CISPR11
Harmonicemissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Class B
ClassA
Comply
Class A with IEC61000-3-2
Complies with IEC61000-3-3
The Pro7 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Guidance and
manufacture’s
declaration
–
electromagnetic immunity
–
for all ME EQUIPMENT and MESYSTEMS
10 of 22
11 of 22
Guidance and
manufacture’s
declaration
–
electromagnetic immunity
The Pro7 is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7 should assure that it is used in such an environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Immunitytest
IEC 60601 testlevel
Compliance level
Electromagnetic environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kVcontact
±
8 kVair
±
6 kVcontact
±
8 kVair
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative humidity
should be at least30%.
Electrical fast
transient/burst
61000-4-4
±
2kV for power
supply lines
±
2kV forpower
supply lines
Mains power quality should be that of
a typical commercial or hospital IEC
environment.
Surge
IEC 61000-4-5
±
1 kV line(s) t
o line(s)
±
1 kVdifferential
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,short
interruptions and
voltage variations
on power supply
input lines
61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0.5cycle
40% U
T
(60%dip in U
T
)
for 5 cycles
<5% U
T
(>95%dip in U
T
)
for 0.5cycle
40% U
T
(60%dip in U
T
)
for 5 cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Pro7
requires continued operation during
power mains interruptions, it is IEC
recommended that the Pro7 be powered
from an uninterruptible power supply ora
battery.
70% U
T
(30%dip in U
T
)
for 25 cycles
70% U
T
(30%dip in U
T
)
for 25 cycles
<5% U
T
(>95%dip in U
T
)
for 5 sec
<5% U
T
(>95%dip in U
T
)
for 5 sec
Power frequency
(50Hz) magnetic
61000-4-8
3A/m
3A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.