2.7
Storage and Transportation
• Store the unit in a place which is not affected by air pressure, temperature (cool),
humidity (dry), ventilation, sunlight, dust, salt, sulfur, etc. for long term storage.
Please be careful not to drop or hit the device during storage or transportation.
• Device function and battery charging should be checked every 2-3 months.
1) Temperature conditions
Condition
Storage
Transportation
Use
Temperature
-20 ~ 60
°
C
-20 ~ 60
°
C
10-35
°
C
2) Humidity conditions
Condition
Storage
Transportation
Use
Percentage
5-90 %RH
5-90 %RH
10-85 %RH
3) Atmospheric pressure
Condition
Storage
Transportation
Use
Pressure
800-1060 hPa
500-1060 hPa
800-1060 hPa
2.8
Periodic Maintenance
Annual maintenance is recommended by a qualified technician for performance and safety, as well as
assurance of accurate X-ray exposure levels.
The battery should be tested and replaced approximately every two years. See section 9 -
Removal and Replacement Procedure
- Medical electrical equipment requires special precautions regarding EMC, and must
be installed and put in to service according to the EMC information provided in the
user manual.
- Portable and mobile RF communication equipment can effect medical electrical
equipment.
- The use of accessories other than those specified in the user manual may result in
increased emissions and void the warranty.