POSDION Co., Ltd.
Page
:
30
Rev.02
PN# 100304-00
REXTAR-X Manual
(2) Ready to X-ray Exposure
< Time to complete preparations: 2 seconds >
After ‘(1) Prepare the X-ray exposure’,
The READY indicator lights up on LCD panel.
(3) X-ray Exposure
< X-RAY Exposure time range:0.01~1.30sec >
Keep pushing Exposure button in ‘(2) Ready to X-ray
Exposure’, X-ray will be expos
ed
w
hen
the X-ray emit
indicator lights up on LCD panel
During Exposure, keep pushing until the X-ray
emission light goes out and the beep sound stops.
If the Exposure button is released before the Exposure is over, Exposure will be immediately
stopped, and Err9 will be displayed on LCD panel. Error 9 is caused when you push X-ray Exposure
release before Exposure setting time.
(4) Standby state after X-ray exposure
< Standby time after exposure: 5sec >
‘(3) X-ray Exposure’ is finished, beep sound will stop
and Standby indicator lights up on LCD panel.
.
※
NOTE
※
If the Exposure button is released before the device goes into
standby mode, the device goes into X-RAY Exposure cancellation
state, and X-RAY Exposure is cancelled.
< Pic1.9.3.2 Ready to X-ray>
< Pic1.9.3.3 X-ray exposure >
< Pic1.9.3.4 Standby State >
and listen for the beeping sound.