2) Exposure time control
Exposure time control is available by push the
increase/decrease buttons.
However, in charging state or during X-ray
exposure not available to control time.
Exposure control time is 0.01 ~ 1.30 sec ( 43 Step ).
( 0.01, 0.02, 0.03, 0.04, 0.05, 0.06, 0.07, 0.08, 0.09, 0.10, 0.12, 0.14, 0.16, 0.18, 0.20, 0.22,
0.24, 0.26, 0.28, 0.30, 0.32, 0.34, 0.36, 0.38, 0.40, 0.42, 0.44, 0.46, 0.48, 0.50, 0.55, 0.60,
0.65, 0.70, 0.75, 0.80, 0.85, 0.90, 0.95, 1.00, 1.10, 1.20, 1.30)
※
NOTICE
※
If you keep pushing time control button, Step moves faster.
*Built-in Safety Trigger
2 step
delay action.
To prevent accidental and unwanted exposures, you will press and hold the
Exposure
button for
1
second
to unlock. Then within 5 seconds, the exposure button must be
pressed and held down continuously to start the exposure cycle. Continue
pressing the
button until the "Beeping" sound stops.
Release of the button too early will cause
an error code to display
and you will need to start the exposure cycle over again.
Remember, once you press the exposure button once to unlock, you have 5 seconds
to again then press and hold to take your image, or it will time out and you will have to
restart the Unlock cycle again.
cycle over again.
P.28
Summary of Contents for REXTAR X
Page 2: ...This page intentionally left blank...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 52: ...POSDION Co Ltd Page 40 Rev 02 PN 100304 00 REXTAR X Manual 3 Block diagram...
Page 63: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 51 This page intentionally left blank...