Page 5
V 1.1 – April 2016
U s e r I n s t r u c t i o n M a n u a l
Installation
Before use, carry out the following steps:
• Visually inspect Ultravap Levante for damage
• Check that correct needle head is supplied and inspect for damage
• Ensure transformer is supplied (if required) and a suitable power cord is available
• Ensure that gas and electrical supplies are correct – see
Specifications & System
Requirements.
• Place Ultravap Levante on a flat surface with a fume hood or well-ventilated area
• Connect the Ultravap Levante to gas and electrical supplies – see
Connecting the Gas &
Power Supply.
• Fit the Needle Head – see
Fitting / Changing Needle Head.
• Carry out adjustments (if needed) of the Ultravap Levante – see
Adjustments.
Report all damaged or missing components to your supplier or distributor immediately.
The Ultravap Levante can be used with 96 well microplates that, as a minimum, conform to the
following standards:
ANSI/SBS 1-2004: Microplates
– Footprint dimensions
ANSI/SBS 4-2004: Microplates
– Well positions
Microplate Requirements