Easy to Attach Setup Sequence:

 Also refer to FM-

1312 (supplied with each Circuit Pack). 

 

1. 

Please review brief instructional setup and usage videos at 
www.porterinstrument.com/silhouette. 

2. 

Use scavenging and set vacuum flow for Silhouette per 
instructions on Page 5 (read Warning on Nitrous Oxide 
exposure minimization on Page 8).   

3. 

Use Disposable Silhouette with Reusable SIL-CONN-KIT 
Connector Kit. Attach smaller diameter hose end to 
attachment barb or flowmeter adapter. Attach larger diameter 
hose end to Porter AVS or other vacuum control device.  

4. 

Place colored sizer masks over the nose to determine 
appropriate mask size for the patient. Sizers may be sterilized 
after each use. 

 

5. 

The Silhouette disposable breathing circuit is packaged in a 
color-coded box and is first removed from the box for 
placement on the patient. 

 

6. 

Prepare for placement of the mask on the patient’s nose by 
removing the adhesive cover tab from the mask. 

 

7. 

Pull the slide bolo down to create a loop large enough to place 
behind the patient’s ears. 

 

8. 

Place nasal barb fully into the right nostril, flexing the mask as 
needed. 

 

9. 

Rotate the mask down over the nose until contact is made. 

 

10. 

Compress the mask adhesive at the bridge of the nose. 

Verify good seal is achieved around the entire mask. 
 

11. 

Place the tubing over the top of the left and right ear. 
 

12. 

  

Move the slide bolo up until the circuit is snug against 

the 

patient’s neck. 

 

13. 

        Connect disposable section union (diameter indexed) to 

gas and vacuum reusable hoses (SIL-CONN-KIT). 
 

14. 

 Observe overall position of the patient, mask, and 

circuit for proper positioning with good mask seal and no evidence 
of tubing kinking.  Use Clip Strap on Reusable Hose to side of 
chair or to patient. 
 

15. 

If End tidal CO2 (EtCO2) 

monitoring is being used, disconnect the 
cap at the end of the disposable breathing 
section and attach the EtCO2 male Luer 
lock sample line. 
 

16. 

You are now ready to turn on your Nitrous Oxide 

flowmeter. 
 
 

 

After patient’s treatment is completed, dispose of the 

Silhouette breathing c

ircuit (36” disposable section only). 

 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 

 
 
Porter Silhouette suggested flow instructions 
 

1. 

Use scavenging and set vacuum flow for 

Silhouette per instructions on Page 5 (read 
Warning on Nitrous Oxide exposure 
minimization on Page 8).  Follow Flowmeter 
directions for use. 

2. 

How much total flow to start

? A 

general determination is the “4-5-6 Rule” where 
6 is the liters per minute (lpm) for an average 
male, 5 is the lpm for the average female, and 4 
is the lpm for a child [values represent Minute 
Breathing Volume] (typical procedure begins 
with 100% O2 and then moves to a chosen 
N2O mixture percentage). 

3. 

Increase flow or percentage as needed 

based on patient observation.   

4. 

Always remind the patient to refrain from 

mouth breathing. 

Summary of Contents for Silhouette-LG

Page 1: ...e Usable with Porter Automatic Vacuum Sw itch AVS 5000 Porter Vacuum Control Block 5501 RK Matrx Vacuum Control Device on flow meter The Quality System for Porter Instrument is Certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of Dental Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use in the ...

Page 2: ...d through to the Vacuum Control Device Disposable Section is disposed after use The sight lines from the medical professional to the patient s mouth work area created by the low profile are superior The medical professional can easily manipulate the patient s mouth without interference to the fit of the Circuit while the patient can turn the head side to side in comfort without losing the seal Use...

Page 3: ...nd of the Reusable vacuum hose 15 to the Union of the Silhouette Disposable Circuit diameter indexed and the other end to the MASK port labeled on body of the AVS 1 Attach a second vacuum hose 15 to the VAC port labeled on body of the AVS 1 then insert straight end of adapter 20 into the other end of the vacuum hose and the tapered end of the adapter into the High Volume Evacuation HVE Line 6 Vacu...

Page 4: ...3 ...

Page 5: ...7a 5068 5 8 T Adapter for In line Vacuum Block See Figure 2 18 5064 Straight Adapter for In line Vacuum Block See Figure 2 19 5065 Vacuum Tube Holder 20 A 3679 000 Adapter Black Round VAC MASK 21 PA 1630 000 Clip Strap 22 SIL SIZER 4 Silhouette Sizers 4 pack Part Numbers Silhouette 4 sizes of Mask Silhouette PD Silhouette SM Silhouette MD Silhouette LG Reorder Part Number SIL PEDO 12 SIL SM 12 SIL...

Page 6: ...gain after adjustment 8 Monitor the vacuum conditions during the procedure by observing the sight glass Note vacuum is indicated with Silhouette Circuit attached and ball pegged high in sight glass Ball float at bottom indicates no vacuum flow Repeat steps 2 through 7 to adjust vacuum flow as necessary 9 Note Adjust vacuum flow with Matrx vacuum control block with Silhouette Circuit attached Set t...

Page 7: ...e mask 11 Place the tubing over the top of the left and right ear 12 Move the slide bolo up until the circuit is snug against the patient s neck 13 Connect disposable section union diameter indexed to gas and vacuum reusable hoses SIL CONN KIT 14 Observe overall position of the patient mask and circuit for proper positioning with good mask seal and no evidence of tubing kinking Use Clip Strap on R...

Page 8: ... for 15 minutes minimum For unwrapped loading Load product into sterilizer without using wrapping packaging or cassette on to standard sterilizer rack or simple sterilizer tray This method of loading is adequate for steam sterilization For wrapped or bag pouch loading Load product into bag or wrap packaging according to sterilizer instructions Assure that the packaging is adequate for steam steril...

Page 9: ...es or 4 not installed by an authorized Parker Hannifin Corporation dealer Parker Hannifin Corporation s obligation for breach of this warranty or for negligence or otherwise shall be strictly and exclusively limited to the repair or replacement of the product or part This warranty shall be void on any product on which the serial number has been altered defaced or removed ORDERS All orders are to b...

Reviews: