Silhouette  is  simple  to  attach  to  a  Flowmeter 
without  a  Bag  Tee.  Please  review  brief 
instructional  setup  and  usage  videos  at 
www.porterinstrument.com/silhouette 

 
 

For Silhouette and Vacuum and Bag Tee:  
(Refer to item (#) in parts list Figure 1) 

 

1.  Existing Bag Tee assembly:  Detach the 3 L 

bag from the bottom of the bag tee (#16). 

 

2.  Fit  the  Silhouette  Cap  to  the  Bag  (bottom) 

port and the Adapter to the 22mm port. 

 

3.  The Circuit design has Disposable and 

Reusable sections (connected to flowmeter 
and vacuum source). Use Silhouette 
Reusable SIL-CONN-KIT Connector Kit.  
Attach smaller ID hose end to flowmeter 
Adapter.  Attach larger ID hose end to AVS 
or other vacuum source. 

 

4.  Attach Vacuum Hoses (#15): Refer to Figure 1. 

 

 

 

 

 

 

 

 

 

 

 

5. 

Automatic Vacuum Switch

:  Attach one end of the 

Reusable vacuum hose (#15) to the Union of the 
Silhouette Disposable Circuit (diameter indexed) 
and the other end to the MASK port (labeled on 
body) of the AVS (#1). Attach a second vacuum 
hose (#15) to the VAC port (labeled on body) of 
the AVS (#1), then insert straight end of adapter 
(#20) into the other end of the vacuum hose and 
the  tapered  end  of  the  adapter  into  the  High 
Volume Evacuation (HVE) Line.  

 

6. 

Vacuum  Control  Block

:  Attach  one  end  of  the 

Reusable  vacuum  hose  (#9)  to  the  Union  of  the 
Silhouette  Disposable  Circuit  (diameter  indexed) 
and the other end to the vacuum control block (#2). 

 

 

The  vacuum  control  block  can  then  be 
inserted  directly  into  the  High  Volume 
Evacuation (HVE) Line; or may be placed “in 
line”  by  cutting  the  vacuum  hose  and 
attaching  the  cut  ends  of  the  tubing  to  both 
ends of the vacuum control block.  NOTE:  To 
properly  read  vacuum  levels,  the  vacuum 
control  block  must  be  held  upright  with  the 
on/off switch above the control valve.   

 
 
 
 
 

 

 

 

 

 
 

 

 

 

 

 

 

 

Summary of Contents for Silhouette-LG

Page 1: ...e Usable with Porter Automatic Vacuum Sw itch AVS 5000 Porter Vacuum Control Block 5501 RK Matrx Vacuum Control Device on flow meter The Quality System for Porter Instrument is Certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of Dental Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use in the ...

Page 2: ...d through to the Vacuum Control Device Disposable Section is disposed after use The sight lines from the medical professional to the patient s mouth work area created by the low profile are superior The medical professional can easily manipulate the patient s mouth without interference to the fit of the Circuit while the patient can turn the head side to side in comfort without losing the seal Use...

Page 3: ...nd of the Reusable vacuum hose 15 to the Union of the Silhouette Disposable Circuit diameter indexed and the other end to the MASK port labeled on body of the AVS 1 Attach a second vacuum hose 15 to the VAC port labeled on body of the AVS 1 then insert straight end of adapter 20 into the other end of the vacuum hose and the tapered end of the adapter into the High Volume Evacuation HVE Line 6 Vacu...

Page 4: ...3 ...

Page 5: ...7a 5068 5 8 T Adapter for In line Vacuum Block See Figure 2 18 5064 Straight Adapter for In line Vacuum Block See Figure 2 19 5065 Vacuum Tube Holder 20 A 3679 000 Adapter Black Round VAC MASK 21 PA 1630 000 Clip Strap 22 SIL SIZER 4 Silhouette Sizers 4 pack Part Numbers Silhouette 4 sizes of Mask Silhouette PD Silhouette SM Silhouette MD Silhouette LG Reorder Part Number SIL PEDO 12 SIL SM 12 SIL...

Page 6: ...gain after adjustment 8 Monitor the vacuum conditions during the procedure by observing the sight glass Note vacuum is indicated with Silhouette Circuit attached and ball pegged high in sight glass Ball float at bottom indicates no vacuum flow Repeat steps 2 through 7 to adjust vacuum flow as necessary 9 Note Adjust vacuum flow with Matrx vacuum control block with Silhouette Circuit attached Set t...

Page 7: ...e mask 11 Place the tubing over the top of the left and right ear 12 Move the slide bolo up until the circuit is snug against the patient s neck 13 Connect disposable section union diameter indexed to gas and vacuum reusable hoses SIL CONN KIT 14 Observe overall position of the patient mask and circuit for proper positioning with good mask seal and no evidence of tubing kinking Use Clip Strap on R...

Page 8: ... for 15 minutes minimum For unwrapped loading Load product into sterilizer without using wrapping packaging or cassette on to standard sterilizer rack or simple sterilizer tray This method of loading is adequate for steam sterilization For wrapped or bag pouch loading Load product into bag or wrap packaging according to sterilizer instructions Assure that the packaging is adequate for steam steril...

Page 9: ...es or 4 not installed by an authorized Parker Hannifin Corporation dealer Parker Hannifin Corporation s obligation for breach of this warranty or for negligence or otherwise shall be strictly and exclusively limited to the repair or replacement of the product or part This warranty shall be void on any product on which the serial number has been altered defaced or removed ORDERS All orders are to b...

Reviews: