the unit from patient service and call the Porter Technical Service Department at 1-888-723-4001. To aid in the 
diagnostics of the unit, make note of the final error code displayed. 
 

Description of Unit:  The DIGITAL MDM mixer consists of a Control Unit attached to a Base Unit via an 
Interface cable.  Note - Control and Base Units are calibrated together and must remain together

 (

match 

serial numbers on each unit). Only use Porter Interface cable (EMI protection).

  The Control Unit interface is 

shown in Figure 1. The device permits the operator to adjust the percent of oxygen through a range of 100% to 
30% (and thereby control the ratio of oxygen to nitrous oxide). The operator can also control the total volume of 
mixed gas up to 9.9 liters per minute.   
 

Mixture Control:  

The DIGITAL MDM mixer precisely mixes oxygen and nitrous oxide from local cylinders or 

pipeline gas systems, and delivers the electronically controlled mixture to the bag tee outlet. 
 

Built-In Safety:

 A series of alarms (see Safety Alarm System page 6) alert the user to any problem condition 

within the DIGITAL MDM mixer unit or in the gas supply. 
 

Displays and Indicators: 

In the following, the location numbers refer to Figure 1. 

 

Flow Display:

 The LPM FLOW display (4) normally displays the total gas flow rate. It also displays the individual 

nitrous oxide or oxygen flow rates when either the N

2

O or O

2

 flow display controls (6) are pressed. The gas flow 

being displayed is indicated in the flow indicator (5) by O

2

, TOTAL or N

2

O. 

 

% OXYGEN Display:

 The % OXYGEN display (3) indicates the oxygen concentration set point in the total flow. 

 

Gas Flow Indicators:

 Gas flow is indicated by bar graph LEDs (1). A green bar graph represents O

2

 while blue 

represents the N

2

O flow. The number of lit bars gives a graphical indication of the flow level of a particular gas. 

 

 

 

Figure 1.  PORTER DIGITAL MDM Mixer Control Unit Front Panel  

(figure shows all LEDs illuminated) 

 

Summary of Contents for DIGITAL MDM

Page 1: ...ration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use by a ph...

Page 2: ...own protection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire...

Page 3: ...A known Protective Earth ground connection must be made via one of the following methods 1 The third wire on the power cord when plugged into an appropriate outlet that is properly installed and grou...

Page 4: ...ropriate regulatory authority in your country state or province Training is recommended to provide a practical hands on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedat...

Page 5: ...or pipeline gas systems and delivers the electronically controlled mixture to the bag tee outlet Built In Safety A series of alarms see Safety Alarm System page 6 alert the user to any problem conditi...

Page 6: ...n restored either by disconnecting reconnecting the power cord or by local power failure the unit will display OFF The operator must press the ON OFF key to restart the unit The unit will then display...

Page 7: ...leasing the key returns the unit to the flow rate and mixture ratio which were in effect prior to the flush Alarm Test Silence Key The alarm test silence key serves two functions When pressed all the...

Page 8: ...8 PORTER DIGITAL MDM SYSTEM ILLUSTRATION...

Page 9: ...alve 9A B 2461 001 N2O Check Valve 10 8005 O2 Hose Diss Diss 5 Ft 10A 8505 N2O Hose Diss Diss 5 Ft 10B 80012 O2 Hose Diss Diss 12 Ft 10C 85012 N2O Hose Diss Diss 12 Ft 11 PA 450 1 002 O2 Hose 12 Ft 11...

Page 10: ...ews again Bracket B 5485 Figure 3 Control Unit mount directly to wall 3 Use bracket shown in Figure 4 for under cabinet mounting Turn same bracket upside down not shown for table top mount Kit B 5555...

Page 11: ...NLET fitting on the back of the mixer head see Figure 6 Hand tighten and then finish tightening with a 7 8 inch open end wrench approximately 1 8 turn Do not over tighten 2 Connect the DISS female hex...

Page 12: ...12 Figure 6 Connection to the back of unit Figure 7 Installation of Power Cord Figure 8 Connections Schematic some components may not be part of your installation Power Cord N2O DISS O2 DISS...

Page 13: ...servoir bag to the bottom downspout of the bag tee Attach the vacuum hoses to the vacuum control source AVS or In line Vacuum Control Block Refer to FM 809 for User Instructions Assembly Installation...

Page 14: ...g the cut ends of the tubing to both ends of the vacuum control block Adjust the vacuum flow using the control knob Vacuum flow is most effective when the ball float is set within the green bar area R...

Page 15: ...ve See Fig 10 below is not functioning properly and should be replaced Figure 10 Emergency Air Valve Check Monthly Place the On Standby switch in the STANDBY Off position Disconnect the corrugated rub...

Page 16: ...1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical p...

Page 17: ...n Failure Test Set the LPM FLOW display to 9 9 LPM Set the OXYGEN display to 50 Turn off the oxygen gas supply to the unit When the gas runs out check for each of the following the OXYGEN FAILURE alar...

Page 18: ...Handbook of Nitrous Oxide and Oxygen Sedation written by Clark and Brunswick and published by Mosby www mosby com This text is a concise and contemporary guide for nitrous oxide oxygen administration...

Page 19: ...nimum Nitrous Oxide Concentration 00 Maximum Nitrous Oxide Concentration 70 Minimum Oxygen Concentration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture...

Reviews: