10 

 

GENERAL INSTALLATION 

 

 

Installation  

1.  Using the available brackets mount the 

Remote Interface Control Unit as desired 
(see Figure 2.) Shown for recessed into wall. 
(Kit B-5555-000) 

 

 

Figure 2.  Control Unit recessed mount. 

 
2.  For mounting on to wall (see Figure 3), first 

remove side screws to separate back bracket 
from Control Unit. Mount back bracket to wall. 
Slide Control Unit on to back bracket and 
install and tighten side screws again. 
(Bracket B-5485) 

 

 

Figure 3. Control Unit mount directly to wall.

 

 
 
 
 

3.  Use bracket shown in Figure 4 for under 

cabinet mounting.  Turn same bracket 
upside-down (not shown) for table-top mount. 
(Kit B-5555-001) 

 

 

Figure 4.  Control Unit under cabinet mount. 
 

4.  Typically mount the Base Unit in a horizontal 

orientation (Figure 5).  Using same mounting 
plate, may also be mounted vertically up, 
down, or side.  

Do not mount Base Unit 

upside-down.

  Adding “Z” brackets allows for 

mounting Base Unit horizontally, but in an 
under cabinet orientation. (Kit B-5555-002, 
“Z” bracket) 

 

 

Figure 5.  Porter Digital MDM connected with 
interface cable. 

Control Unit 

Summary of Contents for DIGITAL MDM

Page 1: ...ration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use by a ph...

Page 2: ...own protection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire...

Page 3: ...A known Protective Earth ground connection must be made via one of the following methods 1 The third wire on the power cord when plugged into an appropriate outlet that is properly installed and grou...

Page 4: ...ropriate regulatory authority in your country state or province Training is recommended to provide a practical hands on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedat...

Page 5: ...or pipeline gas systems and delivers the electronically controlled mixture to the bag tee outlet Built In Safety A series of alarms see Safety Alarm System page 6 alert the user to any problem conditi...

Page 6: ...n restored either by disconnecting reconnecting the power cord or by local power failure the unit will display OFF The operator must press the ON OFF key to restart the unit The unit will then display...

Page 7: ...leasing the key returns the unit to the flow rate and mixture ratio which were in effect prior to the flush Alarm Test Silence Key The alarm test silence key serves two functions When pressed all the...

Page 8: ...8 PORTER DIGITAL MDM SYSTEM ILLUSTRATION...

Page 9: ...alve 9A B 2461 001 N2O Check Valve 10 8005 O2 Hose Diss Diss 5 Ft 10A 8505 N2O Hose Diss Diss 5 Ft 10B 80012 O2 Hose Diss Diss 12 Ft 10C 85012 N2O Hose Diss Diss 12 Ft 11 PA 450 1 002 O2 Hose 12 Ft 11...

Page 10: ...ews again Bracket B 5485 Figure 3 Control Unit mount directly to wall 3 Use bracket shown in Figure 4 for under cabinet mounting Turn same bracket upside down not shown for table top mount Kit B 5555...

Page 11: ...NLET fitting on the back of the mixer head see Figure 6 Hand tighten and then finish tightening with a 7 8 inch open end wrench approximately 1 8 turn Do not over tighten 2 Connect the DISS female hex...

Page 12: ...12 Figure 6 Connection to the back of unit Figure 7 Installation of Power Cord Figure 8 Connections Schematic some components may not be part of your installation Power Cord N2O DISS O2 DISS...

Page 13: ...servoir bag to the bottom downspout of the bag tee Attach the vacuum hoses to the vacuum control source AVS or In line Vacuum Control Block Refer to FM 809 for User Instructions Assembly Installation...

Page 14: ...g the cut ends of the tubing to both ends of the vacuum control block Adjust the vacuum flow using the control knob Vacuum flow is most effective when the ball float is set within the green bar area R...

Page 15: ...ve See Fig 10 below is not functioning properly and should be replaced Figure 10 Emergency Air Valve Check Monthly Place the On Standby switch in the STANDBY Off position Disconnect the corrugated rub...

Page 16: ...1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical p...

Page 17: ...n Failure Test Set the LPM FLOW display to 9 9 LPM Set the OXYGEN display to 50 Turn off the oxygen gas supply to the unit When the gas runs out check for each of the following the OXYGEN FAILURE alar...

Page 18: ...Handbook of Nitrous Oxide and Oxygen Sedation written by Clark and Brunswick and published by Mosby www mosby com This text is a concise and contemporary guide for nitrous oxide oxygen administration...

Page 19: ...nimum Nitrous Oxide Concentration 00 Maximum Nitrous Oxide Concentration 70 Minimum Oxygen Concentration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture...

Reviews: