Type
Default
Parameters
Remark
Appe ndi x A: Met er Fa ct or y Def au lt Set ting
P1.1 Select C/ F
C
Temp unit
P1.2 Nor. Temp
25 C
Selectable: 25 or 20 C
P4.1
P4.2
Viewing
previous
calibration
data
No calibration data for 1st range
No calibration data for 2nd range
P1.3 Temp. Coefficient
2.1%/ C
Adjustable from 0.4 to 10%
P2
.
1
TDS factor
Adjustable from 0.40 to 1.00
(only for 8361)
0.50
P3
.
1
Factory default
Retain your current settings
NO
Appe ndi x B: Cond uc tivi ty to TDS
Conv er si on Fa ct or s
Conductiv ity
at 25 C
TDS KCl
ppm value
Factor
TDS NaCl
Factor
TDS 442
Factor
23
μ
S 11.6 0.5043 10.7 0.4652 14.74 0.6409
84
μ
S 40.38 0.4807 38.04 0.4529 50.5 0.6012
447
μ
S 225.6 0.5047 215.5 0.4822 300 0.6712
1413
μ
S 744.7 0.527 702.1 0.4969 1000 0.7078
1500
μ
S 757.1 0.5047 737.1 0.4914 1050 0.7
2070
μ
S 1045 0.5048 1041 0.5029 1500 0.7246
2764
μ
S 1382 0.5 1414.8 0.5119 2062.7 0.7463
8974
μ
S 5101 0.5685 4487 0.5 7608 0.8478
12,880
μ
S 7447 0.5782 7230 0.5613 11,367 0.8825
15,000
μ
S 8759 0.5839 8532 0.5688 13,455 0.897
80mS
52,168
0.6521 48,384 0.6048 79,688 0.9961
ppm value
ppm value
442: 40% sodium sulfate, 40% sodium
bicarbonate and 20% sodium chloride.
Appendix C: Calculating TDS
conversion factors
The meter can be calibrated by using TDS calibration
standard solutions. The calibration standard requires
the TDS value at a standard temperature such as
25 C. To determine the Conductivity-to-TDS
conversion factor, please use the following formula:
o
Factor = Actual TDS : Actual Conductivity @ 25 C
24