9 When the “CAL” stop flashing, you l
can press” “ less than 1 second
to confirm the value. The meter then
return to conductivity measurement
mode.
If the “CAL” always blinks, please
check the calibration solutions and
make sure it is stable and your input
value in step 8 is correct.
For example 2:
Standard: 1.6mS; Detected :2.1mS
Adjustable range: +0.63ms (2.1*30%)
The values differ less than 30% but
the 1.6mS is already less than
10% range limit (1.99).So, the max.
value could be input is 1.99mS only
Repeat 1~9 for other ranges if needed.
NOTE:
When switch the meter from measure-
ment to calibration mode, the meter
will display the factory default value.
So, if the meter was previously
calibrated, the display may seem to
jump to the factory default value when
entering calibration.
NOTE:
To exit conductivity calibration mode
without confirming calibration, you can
press ” “ ( in step9) more than 2
seconds. This lets you retain the meter's
previous calibration data for the current
range which you proceed.
10.
16