6.Dip the rinsed probe into the other
container. Tap probe on the bottom
of container to remove air bubbles.
Let the probe stabilize to the solution
temperature ( wait about 15 mins)
7.Press “ “ more than2 seconds to
begin calibration. The conductivity
value of solution will blink on LCD.
limit, the displayed
value will equal to the range limit or
10% of range limit.
NOTE:
If the standard buffer is over the
measuring limit or less than 10% of
measuring
8. Press “ “ or “ “ and “ “ to
change the value in order to match
the value to the standard which is
o
referred to normalization temp. 25 C
You can adjust the conductivity
reading up to +30% from the detected
value. However, if your detected value
and standard value differs by more
than +30%, it means cleaning or
replacing meter is needed.
For example:
Standard: 10uS; Detected value:19uS
Adjustable range: +5.7us (19*30%)
However, under above situation, the
values already differed over 30%.
For example 1:
Standard: 22uS; Detected value:19uS
Adjustable range: +5.7us (19*30%)
The values differ less than 30% but the
22uS is already over range limit. So,
the maximum value could be input is
19.99uS only.
15