3-19
3 First Steps
4. Loosen the wing screws that fix the work plate and unfix the locking
hooks.
5. Move the work plate into the front position.
6. Fix the locking hooks and tighten the wing screws again.
7. Fold out the screen again.
3.4
Cabling
In the following you will find a description of all the steps necessary to connect
up the instrument. All the cabling connections are also shown in F
3.18. Secure the connections appropriately.
If you have any problems connecting up the cables, please contact your local
Polytec representative.
WARNING!
Electric shock caused by connection to wrong power supply!
» Connect up the instruments only using a three-pin mains cable to AC
systems 50/60 Hz with a grounded protective conductor, and a nominal
voltage of between 100 V and 240 V.
ATTENTION!
Damage to the instrument caused by overvoltage!
» First of all, connect up all components as described in this chapter before
connecting the mains cable and switching on the instrument.
ATTENTION!
Damage to connections caused by bent contact pins!
All cables must be easy to plug in.
» Check the contact pins before plugging in the cables.
» Do not use any force when plugging in the cables.
» If necessary, replace defective cables with new cables.
» Contact your local Polytec representative.
INFORMATION
The main cable (umbilical) has a plug with a fixed (straight) cable connection
and a connector with a rotatable (slightly angled) cable connection.
You can connect up the rotatable plug either to the front-end or to the
scanning head. As a general rule, we recommend connecting the rotatable
plug to the scanning head. Thus, you avoid bending or twisting the cable and
thereby unnecessarily straining it when moving the scanning head.