Normal Operation 

 

WARNING: Due to the flammable nature of isopentane, the Histology Bath should always be 
used under a fume hood. Be sure to wear appropriate protective equipment, such as splash 
goggles and gloves.  

 

 

NOTE: The Histology Bath operates at maximum cooling performance.  

Place the Power Switch / Circuit Breaker on the front of the Histology Bath in the On position. The 
compressor will start up and the unit will begin cooling. It will take about one hour for the unit to 
achieve a -60°C / -76°F working temperature.  

Once the unit has reached working temperature, remove the reservoir cover and slowly submerge 
the immersion basket in the reservoir. Use the tissue mounts (not supplied) and chuck holder to 
position the samples properly in the sample basket. Replace the reservoir cover.  

 

 

Immersion Basket 

Tissue Mounts 

(user-supplied) 

Chuck Holder 

Summary of Contents for Histology Bath

Page 1: ...Operator s Manual Histology Bath 110 827 09 12 11...

Page 2: ...allation 6 General Site Requirements 6 Adding Liquid to the Reservoir 6 Electrical Power 6 Normal Operation 7 Routine Maintenance Troubleshooting 8 Draining the Reservoir 8 Cleaning the Bath 8 Air Ven...

Page 3: ...ide safe operation Please ensure that all individuals involved in the installation operation or maintenance of this Histology Bath read this manual thoroughly prior to working with the unit This symbo...

Page 4: ...d in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid type that is already in the reserv...

Page 5: ...nty repaired but not replaced If the unit is damaged or does not operate properly contact the transportation company file a damage claim then contact the company where your unit was purchased WARNING...

Page 6: ...110 827 5 Components Controls Power Switch Circuit Breaker Reservoir Cover Drain Valve Drain Port IEC Power Connection Reusable Air Filter behind grille...

Page 7: ...arefully before filling reservoir The Histology Bath is designed for use with isopentane also known as 2 methlybutane Fill bath so that the liquid level is approximately 1 inch 2 54 cm from the top fo...

Page 8: ...on the front of the Histology Bath in the On position The compressor will start up and the unit will begin cooling It will take about one hour for the unit to achieve a 60 C 76 F working temperature O...

Page 9: ...l scouring pads are acceptable The entire unit is housed in a tough well insulated stainless steel casing that is corrosion and chemical resistant Air Vents The unit draws air through the rear and exh...

Page 10: ...ockage Compressor shakes stalls or continually restarts Compressor cycling or overload Check air vents for blockage Check that air flow around unit is not restricted and that room temperature does not...

Page 11: ...nation so as to protect the persons involved in the disposal and recycling of the equipment from health hazards By doing so you will help to conserve natural and environmental resources and you will e...

Page 12: ...the outside of the package A detailed explanation of the defect should be enclosed with the item The warranty shall not apply if the defect or malfunction was caused by accident neglect unreasonable u...

Reviews: