Polimaster PM1701M Operation Manual Download Page 18

 

18 

 

6 CALIBRATION TEST PROCEDURE (REFERENCE) 

 
6.1 General 
This procedure describes the calibration procedure for the PM1701M 

instrument. The calibration tests shall be performed by qualified calibration facilities. 
The instruments shall be calibrated during manufacture, after repair and periodically 
during operation and storage as required by national regulations. 

 
6.2 For operations and measuring instruments refer to the Table below. 

Operation 

Paragraph 

Reference and auxiliary measuring 

instruments  

External 

examination 

6.6.1 

Check-up 

6.6.2 

Testing of 

metrological 

characteristics 

6.6.3.1, 

6.6.3.2 

Dosimeter calibration assembly with a 
reference source

  137

Cs in compliance with 

ISO 4037-1. Certified uncertainty of the 
assembly not more than 5 % at 0.95 
confidence probability 

6.4 

Barometer. Least division 1 kPa. 
Measurement range from 60 to 120 kPa 

6.4 

Thermometer. Least division 0.1 

°

C. 

Measurement range from 10 to 30 

°

6.4 

Hygrometer. Measurement range from 30 to 
90 % 

6.6.3.1, 

6.6.3.2 

Stopwatch. Measurement range from 1 to 
600 s 

6.4 

Dosimeter. Intrinsic error 

±

 15 %. 

 

6.3 Safety precautions  
During the calibration procedure observe the existing national standards for 

handling radioactive materials  and other sources of ionizing radiation and radiation 
safety regulations, as well as on-site accident prevention instructions. 

 
6.4 Test conditions: 
- ambient temperature 

 

 

 

(20 

±

 5) 

°

C; 

- relative air humidity 

 

 

 

(60 

±

 15) %; 

- atmospheric pressure    

 

 

(101,3 

±

 4) kPa; 

- background gamma radiation:  

 

not more than 0.20 

µ

Sv/h. 

 
6.5 Preparation for calibration tests 
Prior to calibration, study this Operating Manual and prepare the instrument 

according to the Operating Manual. 
 

Summary of Contents for PM1701M

Page 1: ...PM1701M ALARMING RATEMETER Operation Manual...

Page 2: ...Modes 10 2 2 1 Self Check 10 2 2 2 Background Calibration 10 2 2 3 Search Mode Detection and Location of Radiation Sources 11 2 2 4 Settings 12 2 2 4 1 Setting the Coefficient n 12 2 2 4 2 Switching O...

Page 3: ...nstrument alerts the user by visual and audible signals about the presence of radiation source nearby All operation history is stored in the instrument s non volatile memory and may be downloaded to a...

Page 4: ...e intrinsic error of count rate measurement temperature and humidity variations from normal to high temperature variations from normal to low 15 o C extreme values of supply voltage 40 10 10 PC commun...

Page 5: ...the instrument off and follow instructions above to replace the battery 1 4 Design Control Buttons and Display The instrument Fig 1 is composed of measuring unit 1 attached to telescopic pole 4 The ma...

Page 6: ...6 Figure 1 PM1701M The control buttons view A are used to 12 MODE Button turn the instrument on update background press and hold the button until the LCD displays CAL message...

Page 7: ...itch off Vibr on headphones alarm in the setting mode LCD indication 15 LCD upper line count rate in count per second s 1 messages test CAL Aud etc coefficient n value in the setting mode Analog Scale...

Page 8: ...ight alarm The 16 bit RISC based processor is used to test of the instrument upon power on process data control all operating modes display processed information on LCD control audible alarms monitor...

Page 9: ...signals increases as the device moves closer to a radiation source The audible alarms can be switched on off in the PC communication mode or using the control buttons on the front panel of the instrum...

Page 10: ...ally occurring background radiation The processor counts pulses coming from the detector for the preset time and the analog scale segments increase as the instrument calibrates The scale completely fi...

Page 11: ...in the processor memory are updated every 0 25 s Every 0 25 second the current average counts are compared with the alarm threshold calculated during background calibration If the current average cou...

Page 12: ...dating the background Keeping the same distance to the scanned object press and hold the MODE button 12 or button 10 Fig 1 if the headphones is used until CAL is displayed to start the calibration seq...

Page 13: ...m with the control buttons provided that this mode is enabled in the parameters set in the PC communication mode If this mode is enabled the built in audible alarm is switched on off in the following...

Page 14: ...omputer Install the Software PM17 _14XX supplied on CD Run SETUP EXE program using the installation instructions install doc autorun Help file and attached text document describe how to use the progra...

Page 15: ...on off audible alarms enable disable switching on off the alarm using control buttons check or change coefficient n check or change the time and date check or change the logging intervals in which co...

Page 16: ...ery replacement and regular check up of the instrument Routinely examine the instrument body remove any dust and decontaminate the instrument in situation of direct contact with radioactivity Wipe the...

Page 17: ...st of the problems can be traced to low battery or poor quality battery Use only fresh high quality batteries There are no user serviceable parts in the instrument If repairs are required contact the...

Page 18: ...f the assembly not more than 5 at 0 95 confidence probability 6 4 Barometer Least division 1 kPa Measurement range from 60 to 120 kPa 6 4 Thermometer Least division 0 1 C Measurement range from 10 to...

Page 19: ...wait at least 60 s Then take 5 background readings Nbi s 1 and calculate b N s 1 by formula 5 1 5 1 i bi b N N 6 1 2 place the instrument on the dosimetric calibration assembly with 137 Cs gamma sour...

Page 20: ...20 Compare main relative error to limits of permissible main error acc 35 If 35 the instrument is rejected as defective if 35 the instrument is considered ready for use...

Page 21: ...r instruments should be packaged in hermetic compartments The instruments should be stored in the original manufacturer s package with the air temperature between 15 to 50 C and humidity less than 95...

Page 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the instrument 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authori...

Reviews: