Polimaster PM1701M Operation Manual Download Page 10

 

10 

2.2 Operating Modes 

 

 

Self-Check (Test) 

 

Background Calibration  

 

Search Mode (count rate indication, s

-1

 

Settings 

 

PC Communication via IR Interface 

 

 
2.2.1 Self-Check  

The instrument goes into this mode when turned on. Before 

the self-tests,  an  audible  alarm is shortly  activated, then all 
segments and indicators appear on the LCD. The instrument 
displays the

 

“tESt” message and goes through the self-test 

sequence: LCD test, detector test, processor test, non-volatile 

memory test, alarm test. The analog scale segments count down the test time. After 
the self-check, the instrument  shortly produces an audible signal, then goes into the 
background calibration. The LCD displays "CAL" message and the analog scale with 
increasing segments.

 

 
2.2.2 Background Calibration 

The instrument goes into this mode automatically following 

completion of self-check, "CAL" message appears on the LCD. In 
the Background Calibration mode the instrument measures 
naturally occurring background radiation. The processor counts 

pulses coming from the detector for the preset time, and the analog scale segments 
increase as the instrument calibrates. The scale completely filled with segments 
indicates that the background calibration is complete. Later on when the user makes 
the instrument update the background, the background count time will automatically 
decrease with increasing the background level. 

 
The processor calculates the alarm threshold value:   
 

N

n

N

Alarm

+

=

 

 

where 

N

 – the average count time for the calibration time;

 

n

 

 the number of meansquare deviations (coefficient 

n

 
The coefficient 

n

 changes the alarm threshold: the smaller the value of 

n

, the 

smaller the threshold value, and the greater the instrument’s sensitivity

However, 

this also increases the probability of false alarms.  

The user may run background update whenever it is necessary during normal 

operation of the instrument. To do so, press and hold (for more than 2 seconds) the 
MODE button 12 (Fig.1) until "CAL" and the analog scale with increasing segments 
appears on the display. The instrument is acquiring new background. With connected 

Summary of Contents for PM1701M

Page 1: ...PM1701M ALARMING RATEMETER Operation Manual...

Page 2: ...Modes 10 2 2 1 Self Check 10 2 2 2 Background Calibration 10 2 2 3 Search Mode Detection and Location of Radiation Sources 11 2 2 4 Settings 12 2 2 4 1 Setting the Coefficient n 12 2 2 4 2 Switching O...

Page 3: ...nstrument alerts the user by visual and audible signals about the presence of radiation source nearby All operation history is stored in the instrument s non volatile memory and may be downloaded to a...

Page 4: ...e intrinsic error of count rate measurement temperature and humidity variations from normal to high temperature variations from normal to low 15 o C extreme values of supply voltage 40 10 10 PC commun...

Page 5: ...the instrument off and follow instructions above to replace the battery 1 4 Design Control Buttons and Display The instrument Fig 1 is composed of measuring unit 1 attached to telescopic pole 4 The ma...

Page 6: ...6 Figure 1 PM1701M The control buttons view A are used to 12 MODE Button turn the instrument on update background press and hold the button until the LCD displays CAL message...

Page 7: ...itch off Vibr on headphones alarm in the setting mode LCD indication 15 LCD upper line count rate in count per second s 1 messages test CAL Aud etc coefficient n value in the setting mode Analog Scale...

Page 8: ...ight alarm The 16 bit RISC based processor is used to test of the instrument upon power on process data control all operating modes display processed information on LCD control audible alarms monitor...

Page 9: ...signals increases as the device moves closer to a radiation source The audible alarms can be switched on off in the PC communication mode or using the control buttons on the front panel of the instrum...

Page 10: ...ally occurring background radiation The processor counts pulses coming from the detector for the preset time and the analog scale segments increase as the instrument calibrates The scale completely fi...

Page 11: ...in the processor memory are updated every 0 25 s Every 0 25 second the current average counts are compared with the alarm threshold calculated during background calibration If the current average cou...

Page 12: ...dating the background Keeping the same distance to the scanned object press and hold the MODE button 12 or button 10 Fig 1 if the headphones is used until CAL is displayed to start the calibration seq...

Page 13: ...m with the control buttons provided that this mode is enabled in the parameters set in the PC communication mode If this mode is enabled the built in audible alarm is switched on off in the following...

Page 14: ...omputer Install the Software PM17 _14XX supplied on CD Run SETUP EXE program using the installation instructions install doc autorun Help file and attached text document describe how to use the progra...

Page 15: ...on off audible alarms enable disable switching on off the alarm using control buttons check or change coefficient n check or change the time and date check or change the logging intervals in which co...

Page 16: ...ery replacement and regular check up of the instrument Routinely examine the instrument body remove any dust and decontaminate the instrument in situation of direct contact with radioactivity Wipe the...

Page 17: ...st of the problems can be traced to low battery or poor quality battery Use only fresh high quality batteries There are no user serviceable parts in the instrument If repairs are required contact the...

Page 18: ...f the assembly not more than 5 at 0 95 confidence probability 6 4 Barometer Least division 1 kPa Measurement range from 60 to 120 kPa 6 4 Thermometer Least division 0 1 C Measurement range from 10 to...

Page 19: ...wait at least 60 s Then take 5 background readings Nbi s 1 and calculate b N s 1 by formula 5 1 5 1 i bi b N N 6 1 2 place the instrument on the dosimetric calibration assembly with 137 Cs gamma sour...

Page 20: ...20 Compare main relative error to limits of permissible main error acc 35 If 35 the instrument is rejected as defective if 35 the instrument is considered ready for use...

Page 21: ...r instruments should be packaged in hermetic compartments The instruments should be stored in the original manufacturer s package with the air temperature between 15 to 50 C and humidity less than 95...

Page 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the instrument 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authori...

Reviews: