Deep C Meter 300 AD™
operation manual
16
SMB-001 Deep C Meter 300 AD Manual
Figure 7
Cell to Zn coupon
ROV-II™ calibration wiring schematic
Calibration
voltmeter display
at 2 Volts DC setting
Seawater solution
Non-metallic bucket
Step
+VE
—VE
DC scale
Expected result
01
Tip / Pin 1
Reference electrode 2 / Pin 3
2 V
–1.030 to –1.070
02
Tip / Pin 1
Reference electrode 1 / Pin 2
2 V
–1.030 to –1.070
Zn calibration coupon
(Readings should
be within 0.005V
of each other)