Polatrak Deep C Meter 300 AD Operation Manual Download Page 10

www.polatrak.com

   |   

13813 FM 529 Rd, Houston, TX 77041 USA

   |   

Telephone: +1 (713) 983 7117

   |   

Email: [email protected]

©

 2015 Deepwater Corrosion Services Inc. Specifications subject to change without notice

Deep C Meter 300 AD™

 operation manual

10

SMB-001 Deep C Meter 300 AD Manual

8.1.19

  Using a 7/16” (or 10mm) wrench, ensure that the contact tip is tight on the nose cone.

8.1.20

  Ensure that the tip is sharp.

 

9.  Taking CP potential readings

9.1.

  Stab the contact probe onto the point on the structure where the reading is required; when a 

good contact is made, the Deep C Meter 300 AD™ will show a steady reading. 

CAUTIONS

•  Avoid stabbing painted or coated surfaces; whenever possible, try to take readings on bare steel or 

on spots specially designated for CP measurement. CP stab points are typically labeled with “CP”.
•  Anodes can be stabbed if necessary; however, it is always best to stab the anode band where 

possible, as it is typically grounded to the structure.
•  Avoid trying to stab through heavy, hard marine growth if possible.
•  The probe unit should be kept out of the mud to prevent contamination of the electrodes.
•  If the readings are not steady or do not match the criteria in Table 3 below, the most likely cause 

is a high-resistance contact. DO NOT RECORD THESE NUMBERS. Continue to stab until readings 

stabilize and follow the recommended action. See troubleshooting section 13 if problems persist.
•  NEVER immerse the system with any connectors un-mated unless the dummy plugs are installed.

10.  Data interpretation

10.1  

For carbon steel structures in seawater, the readings obtained should be within the ranges 

specified in Table 3. Exceptions may occur if the probe is stabbed onto an isolated section of a 

corrosion-resistant alloy such as stainless steel, copper or nickel. In these cases, Table 3 no longer 

applies and readings less negative than indicated in the table may be noted. If this is the case, en-

sure that the readings are steady and the electrodes maintain the same potential difference between 

each other during the last calibration check.

Table 3: Normal cathodic protection ranges for bare carbon steel in seawater

Range (mV)

Interpretation

Action

-500 or more positive

Error

Remake contact & verify  

Measure two other points around the 

component 

Abort & refer to troubleshooting section

-501 to -649

Isolated from cathodic protection

Remake contact & verify 

Record data

-650 to -799

Not cathodically protected

-800 to -849

Marginal cathodic protection

-850 to -1049

Cathodically protected

Record data

-1050 to -1149

Anode potential

-1150 or more negative

Error

Remake contact & verify 

Measure two other points around the 

component 

Abort & refer to troubleshooting section

10.2  Removal from ROV

10.2.1

  When equipment is removed from the ROV, wash thoroughly with fresh water and allow to 

dry. Soak the ROV-II and EFG probe units in fresh water for at least one hour. Replace lens cover 

on Deep C Meter 300 AD™ and place it in the protective case provided.

Summary of Contents for Deep C Meter 300 AD

Page 1: ...B 001 Deep C Meter 300 AD Manual 2015 Deepwater Corrosion Services Inc Specifications subject to change without notice Deep C Meter 300 AD ROV mounted CP probe Operation manual ROV interfaced cathodic protection monitoring system Can output both contact and EFG probe readings Digital output RS232 standard ASCII string No batteries ROV 24V DC powered ...

Page 2: ...Deep C Meter AD calibration 8 7 2 Bucket calibration 9 7 3 Subsea calibrations 9 7 4 Online calibration 9 8 Operation 10 8 1 Mounting on ROV 10 9 Taking CP potential readings 11 10 Data interpretation 11 10 1 Removal from ROV 11 11 Maintenance and repair 12 11 1 Tip replacement 12 11 2 Electrode element replacement 12 12 RS232 functions 13 12 1 Data output 13 12 2 Control data input 13 13 Troubles...

Page 3: ...able 7 List of spare parts 15 1 Overview The Deep C Meter 300 AD is a complete ROV interfaced cathodic protection monitoring system The system comprises five basic parts 1 1 LED readout Triple voltmeters housed in a Delrin pressure housing 1 2 Contact probe Twin element tip contact CP probe Polatrak Model ROV II 1 3 Flying lead Three meter long umbilical that connects the probe to the readout 1 4 ...

Page 4: ...sure housing to change batteries 2 Documentation The following documents are contained in the appendix 2 1 ROV II Operations Manual 354 MN01 ENG 2 2 EFG 1 Operations Manual 358 MN02 ENG 2 3 RUSS Operation Manual 358 MN04 ENG 2 4 Parker Seals Parker O Lube MSDS O ring lubricant 2 5 Dow Corning 4 Electrical Insulating Compound MSDS connector sealant PLEASE CONTACT YOUR DEEPWATER REPRESENTATIVE FOR A...

Page 5: ...he inside of the pressure housing is gained by pulling the lens Removing the lens in the field should be avoided wherever possible 4 1 3 Bulkhead connector The back end of the pressure housing is sealed with three bulkhead connectors These connectors thread into the back of the pressure housing and also seal with o rings It should not be necessary to remove these connectors during normal maintenan...

Page 6: ...C Meter 300 AD to the ROV frame and provide angled adjustment to optimize camera viewing of the displays The mount and all fixing hardware are made of Delrin 4 3 Flying leads 4 3 1 There are three leads supplied per kit each serving a unique function 4 3 2 A full wiring diagram can be found in section 6 0 Figure 4 Meter mount Figure 5 Flying leads ROV II meter side Female 3 pin EFG probe side Male...

Page 7: ...placeable contact tip it also includes the female connector that attaches to the male pin on the extension cable from the tail unit Tips are made from stainless steel and plated with nickel the nose cone is fabricated from a rugged thermoplastic 5 4 Reference electrode element Two plug in silver silver chloride elements are included Electrode elements are accurate to 5 mV See the operations and ma...

Page 8: ...he probe tip The upper two display lines on the LED readout should read between 1 030V and 1 070V and be within 0 005V of each other NOTE If the probe has not been used in the past 24 hours it may take 15 30 minutes for the probe to reach equilibrium EFG calibration check only applicable if using the EFG probe 7 2 4 Ensure both sets of holes on the EFG probe are submerged Ensure there is no trappe...

Page 9: ... 8 1 4 If using zoom cameras position the camera approximately 12 away from the edge of the lens 8 1 5 If using fixed zoom cameras physically position the camera approximately 0 5 to 3 away from the edge of the lens 8 1 6 Ensure that all fasteners are tightly secured and in good condition Verify that all lock nuts and washers are in place on the readout unit mount 8 1 7 Connect and route the power...

Page 10: ...f problems persist NEVER immerse the system with any connectors un mated unless the dummy plugs are installed 10 Data interpretation 10 1 For carbon steel structures in seawater the readings obtained should be within the ranges specified in Table 3 Exceptions may occur if the probe is stabbed onto an isolated section of a corrosion resistant alloy such as stainless steel copper or nickel In these ...

Page 11: ...ltmeter set on 2 00 VDC range stab the large pin socket pin 1 with the positive voltmeter lead and stab pin socket 2 first clockwise of large pin with the negative lead note reading Then repeat stabbing pin socket 3 with the negative lead and note the reading The faulty electrode will be the one that gives the lowest least negative reading 11 2 6 Ensure that the probe is drained of seawater 11 2 7...

Page 12: ...300 AD operation manual 12 SMB 001 Deep C Meter 300 AD Manual 12 RS232 Functions 12 1 Data output 12 1 1 Data string structure Stab 1 Stab 2 EFG For example 1034 1035 12 2 1034 1036 12 2 12 1 2 After each data set there is a line feed and carriage return present The rate is 2 data sets per second Table 4 RS232 communication settings Speed Baud 9600 Data bits 8 Stop bits 1 Parity None Flow control ...

Page 13: ...n real time Comma separator Command SD Tab separator Command SDHT Analog polarity command SP polarity Sets the polarity of analog voltages being measured by the ADC There is no response for this protocol Positive character sets the common analog ground to the negative side of the converter Positive Command SP Negative Command SP Online calibration Cal Re calibrates the internal AD which takes appr...

Page 14: ...flying lead is nicked Inspect and repair replace as necessary Displays are blank Whip is cut or power polarity is inverted Check cabling and wiring Displays are too dim or blurry Intensity is set too high or low for the available camera Increase intensity of light shining on meter Adjust the intensity using the command line See Table 5 RS232 Command List Garbled data stream output Ground fault or ...

Page 15: ...0009 Lens O ring 2 4 GSK0008 Lens backup ring 1 2 FAS0107 Lens retaining screw 4 1 MFR0016 Lens cover 1 0 UWC0038 Power bulkhead dummy plug female 1 0 UWC0044 EFG bulkhead dummy plug male 1 0 UWC0045 ROV II bulkhead dummy plug female 1 0 UWC0041 ROV II flying lead 1 0 UWC0035 Power cable whip 1 0 UWC0061 EFG flying lead Optional Optional MLT0042 EFG probe Optional Optional MLT0026 ROV II probe nos...

Page 16: ...manual 16 SMB 001 Deep C Meter 300 AD Manual Figure 7 Cell to Zn coupon ROV II calibration wiring schematic Calibration voltmeter display at 2 Volts DC setting Seawater solution Non metallic bucket Step VE VE DC scale Expected result 01 Tip Pin 1 Reference electrode 2 Pin 3 2 V 1 030 to 1 070 02 Tip Pin 1 Reference electrode 1 Pin 2 2 V 1 030 to 1 070 Zn calibration coupon Readings should be withi...

Reviews: