6 ControlBox indicator light
Planmeca ProSensor ControlBox indicator light explanation
CONTROLBOX INDICATOR LIGHT
PLANMECA PROSENSOR STATUS
Off
Planmeca ProSensor system power off
Dim blue
Planmeca ProSensor system is off (not in
intraoral exposure-mode and the cable is
connected to the ControlBox)
Bright blue
Planmeca ProSensor system is on (Imaging
program communicates with the Planmeca
ProSensor system)
Slowly flashing blue
Waiting for Ready
Steady green
Waiting for Exposure
Rapidly flashing green
The exposure is taken and image is transferred
from the sensor to the ControlBox
Steady red
Error mode
Slowly flashing yellow
Service mode
Uploading ControlBox software
Slowly flashing blue, turns to slowly flashing dim
blue, then to quickly flashing dim blue
Reading calibration files from the sensor
Flashing violet
ControlBox startup with back-up software
Flashing white
Sensor is being calibrated
Purple light at beginning when powering up the
control box
Control box is running a factory software
NOTE
The exposure can only be taken when the Planmeca ProSensor ControlBox
indicator light is green and steady, not when the indicator light is flashing.