14 Exposure value tables for Planmeca ProX
14.1 Default exposure values
The following table shows the default exposure values for Planmeca
ProSensor with no target selected.
Short cone
Long cone
kV
mA
s
kV
mA
s
Large
63
8
0,1
Large
63
8
0,2
Small
60
8
0,08
Small
60
8
0,16
14.2 Recommended exposure values for animals with Planmeca ProX
Default exposure values described in A.1 are factory settings, by changing
only the time of these default exposure values you can easily set and save
the recommended exposure values for animals.
For detailed instructions on how to program new exposure values for each
ProX exposure region, see Planmeca ProX for veterinary use user's manual.
Cats or smaller animals
Larger animals
kV mA s
kV mA s
Incisors
Upper jaw
60 8
0.100
Incisors
Upper jaw
60 8
0.160
Lower jaw
60 8
0.100
Lower jaw
60 8
0.160
Premolars Upper jaw
60 8
0.100
Premolars Upper jaw
63 8
0.200
Lower jaw
60 8
0.100
Lower jaw
63 8
0.200
Molars
Upper jaw
60 8
0.100 –
0.125
Molars
Upper jaw
63 8
0.250
Lower jaw
60 8
0.100 –
0.125
Lower jaw
63 8
0.250
Occlusal
Upper jaw
66 8
0.125
Occlusal
Upper jaw
70 8
0.250
Lower jaw
66 8
0.100
Lower jaw
70 8
0.200
Needle in
root canal
Depends on
teeth
60 8
0.100 -
0.125
Needle in
root canal
Depends on
teeth
60 8
0.160 -
0.250
Bitewing
Not used for animals
Bitewing
Not used for animals
14 Exposure value tables for Planmeca ProX
User's manual
Planmeca ProSensor® HD 23