1 Incisors
Exposure kV/mA: 60 kV, 8 mA
Exposure time:
Upper jaw 0.100 s
Lower jaw 0.100 s
2 Premolars
Exposure kV/mA: 60 kV, 8 mA
Exposure time:
Upper jaw 0.100 s
Lower jaw 0.100 s
3 Molars
Exposure kV/mA: 60 kV, 8 mA
Exposure time:
Upper jaw 0.100 s - 0.125 s
Lower jaw 0.100 s - 0.125 s
4 Occlusal
Exposure kV/mA: 66 kV, 8 mA
Exposure time:
Upper jaw 0.125 s
Lower jaw 0.100 s
5 Needle in root
canal
Exposure kV/mA: 60 kV, 8 mA
Exposure time: 0.100 s - 0.125 s
Note, fine tune the exposure parameters depending
on teeth.
5 Bitewing (BW) Not used for animals (or use the same parameters
as for molars).
7.2.2 Recommended exposure values for larger animals
The following table shows the recommended exposure values for Planmeca
ProSensor.
NOTE
These values are for digital sensor.
7 Before exposure
User's manual
Planmeca ProSensor® HD 9
Summary of Contents for ProSensor HD
Page 1: ...PlanmecaProSensor HD for veterinary use user s manual EN 30028714...
Page 4: ...Table of contents Planmeca ProSensor HD User s manual...
Page 28: ......
Page 29: ......