- 49 -
Move the fluorescent screen so that it is going from left and right to up and down,
respectively.
NOTE: The touch screen should still be in
Primary Collimator
showing the
d
collimator and the sensor
alignment tool should still be on the sensor holder.
Press the exposure switch to see where the beaver tall is collimating the beam. It
should fill both the top and bottom within the box on both fluorescent screens.
(Figure 88)
U P
D O W N
Figure 88