- 13 -
NOTE: Make sure beam alignment tool is attached to the sensor alignment tool. (Figure 13)
Figure 13
1.1 Collimator 0 (adult)
Now press the exposure button to see where the radiation beam is hitting the
screen.
The beam image should appear within the borders of the rectangle markings
on the alignment tool. (Figure 14)
U P
D O W N
Figure 14
If not too left, right, up, down, or angled see below.