Ball Phantom
4.1 Setting up Touch Screen
Place the ball phantom in the Patient Positioning Mechanism. (Figure 36)
Figure 36
Alignment lasers
Hold the layer right button for 1 second. (Figure 37)
Figure 37
Adjust the layer so that it is at zero with the layer arrow fields. (Figure 38)
Figure 38