49
Additional information
3. The screen will show that data can now be written via the network.
4. Press
OK
to return to read-only mode.
5.3
Specifications
5.3.1
System
Dimensions
435 mm wide x 330 mm deep x 185 mm high
Weight
17 kg
Storage temperature
-10 °C to +50 °C
Storage humidity
5% to 95% relative humidity non-condensing
Storage special instructions
Recharge every 4 months by connecting to the mains
power supply for 24 hours.
Operating environment
For indoor use only
Operating temperature
+5 °C to +40 °C for safe operation
See
Control
table for control limitations.
Operating humidity
20 % to 80 % relative humidity non-condensing
decreasing linearly to 50 % relative humidity at 40 °
C.
Altitude
up to 2000 m
Pollution degree
Pollution degree 2 (BS EN61010-1)
IP rating
IP31
5.3.2
Control
Temperature control range
(a 5 °C) to (a 20 °C)
40 °C max.
Temperature measurement
accuracy
± 0.2 °C
Temperature control accuracy
± 0.1 °C measured after any transient effects due to
set-point changes have subsided.
Flow control range
0 ml/minute to 900 mL/minute
Normalised to 0 C , 50% RH and 1 bar.
Flow accuracy
The greater of ± 10% or ± 3 ml/minute
Flow control accuracy
The greater of ± 5% or ± 2 ml/minute measured after
any transient effects due to set-point changes have
subsided.
Accuracies apply at the calibration points.
The system is factory calibrated for an operating temperature of 37°C, nominal bleed
flow of 30 mL/min and a purge at 360mL/min.
49
Summary of Contents for BT37-02
Page 1: ...INSTRUCTIONS FOR USE BT37 v2 MA200214 M EN 2 0 24...
Page 2: ......
Page 5: ...Introduction...
Page 17: ...Installation...
Page 21: ...Operation...
Page 32: ......
Page 33: ...Routine maintenance and troubleshooting...
Page 46: ......
Page 47: ...Additional information...
Page 55: ...55 Index view side 12 W warnings 8 warranty 6...
Page 56: ......