36
Routine maintenance and troubleshooting
6. Visually inspect the surface to ensure that all visible soil has been removed.
Disinfection
1. Prior to disinfection, the incubator chamber must first be cleaned by following the
cleaning procedure above.
2. Spray the surface with isopropyl alcohol at 70% v/v dilution.
3. Allow to soak for 15 minutes at room temperature.
4. Remove the disinfectant with a clean non-linting cloth (gauze). Use cotton buds or
swabs where necessary to ensure contact is made with all grooves and corners of
the surface plate.
5. Repeat steps 2, 3 and 4 one more time.
6. Wipe the surface over with sterile water and a clean non-lint cloth to remove any
residual fluids. Use cotton buds or swabs where necessary to ensure contact is
made with all grooves and corners of the surface plate.
7. Leave the unit to dry until all residual cleaning fluids have evaporated.
4.4
Checking the liquid level indicator
1. Look through the liquid level indicator and ensure bubbles can be seen.
4.5
Checking the battery
1. Ensure the BT37-02 has been running for at least 24 hours.
2. In normal run mode, disconnect power.
3. Acknowledge the power fail alarm.
4. Confirm the unit can run from the battery for 30 minutes.
5. Reconnect the mains supply.
Summary of Contents for BT37-02
Page 1: ...INSTRUCTIONS FOR USE BT37 v2 MA200214 M EN 2 0 24...
Page 2: ......
Page 5: ...Introduction...
Page 17: ...Installation...
Page 21: ...Operation...
Page 32: ......
Page 33: ...Routine maintenance and troubleshooting...
Page 46: ......
Page 47: ...Additional information...
Page 55: ...55 Index view side 12 W warnings 8 warranty 6...
Page 56: ......