PL Engineering Ltd
Z-Meter DX4091/ User Guide
page 26 of 44
ver.1.02 (2022)
Apply a small amount of low-
melting tin solder on one end side
of the thermoelement using a
soldering iron (and under a
microscope is preferable).
Place the sample at the wire
crosshairs on the frame (between
wires) side with solder on top.
Carefully solder the upper side of
the sample to the upper wire with
a soldering iron.
Flip the test frame.
Using a small amount of solder
and flux, solder the second side
of the sample to the second wire.