2.4. Design overview
The Z-Meter is a single-channel device with internal chamber for
placement of test frame with mounted pellet sample. Chamber’s
cover is magnetically locked when closed or opened.
The internal chamber is a passive thermostat which provides
constant temperature during measurements. Temperature inside the
chamber is measured by integrated temperature sensor.
Spot heater inside the chamber is placed for measurement of pellet
conductivity type.
There are also LEDs for visual control of measurements status.
Cable with USB connector for communication with PC.
2.5. Features
-
The device provides measurement of performance parameters:
pellet resistance
R
, electric conductivity
𝜎𝜎
and Figure-of-Merit
Z
.
-
Measured Figure-of-Merit (
Z
) allows calculation of cooling
performance of TE modules (
Δ
T
max
) to be made with use of such
pellets.
-
Resistance (
R
) of the pellet also allows calculating resistance of
TE module (
ACR
) to be made with use of such pellets with
known number (
N
) of pellets (
ACR
≈N x R
).
-
The Z-Meter allows checking of type of pellets conductivity (n-
type or p-type).
Summary of Contents for Z-Meter DX4091
Page 1: ...Moscow 2022 version 1 02 Z METER DX4091 USER GUIDE PL Engineering Ltd...
Page 4: ...PL Engineering Ltd Z Meter DX4091 User Guide page 4 of 44 ver 1 02 2022...
Page 6: ...PL Engineering Ltd Z Meter DX4091 User Guide page 6 of 44 ver 1 02 2022...
Page 41: ...Z Meter DX4091 User Guide PL Engineering Ltd ver 1 02 2022 page 41 of 44...
Page 42: ...PL Engineering Ltd Z Meter DX4091 User Guide page 42 of 44 ver 1 02 2022...
Page 43: ...Z Meter DX4091 User Guide PL Engineering Ltd ver 1 02 2022 page 43 of 44...