![Pike MIRacle Installation And User Manual Download Page 9](http://html1.mh-extra.com/html/pike/miracle/miracle_installation-and-user-manual_1555976009.webp)
PN 350-007660-07
P a g e
|
5
Note:
When using the Digital Force Adapter in conjunction with the Three-Reflection ATR MIRacle
plate, the 7.8 mm nylon press tip must be used. The High-Pressure Clamp will damage the crystal if
used with the metal press tip.
6.
Insert the connector of the load cell assembly and the power cord into the back of the LCD monitor.
The appropriate location of each connector is clearly labeled. If the load cell is not connected to the
LCD monitor prior to applying power, erroneous readings are displayed.
7.
Choose an appropriate electrical outlet adapter and attach it to the power cord. Plug the power cord
into the electrical outlet.
8.
Within the first 24 hours it is normal to experience drift. Therefore, 24 hours stabilization time prior
to use is recommended after initially connecting the LCD monitor into an electrical outlet. Due to
the long stabilization time, leaving the Digital Force Adapter assembly powered up is recommended.
Installation for GladiATR or GladiATR Vision
1.
Remove the High-Pressure Clamp from the GladiATR base optics.