PN 350-007660-07
P a g e
|
6
2.
With the GladiATR clamp removed from the base, remove the press tip end from the High-Pressure
Clamp by turning the tip counter clockwise.
3.
Attach the load cell assembly by screwing it onto the small threaded rod that held the press tip until
finger tight. Exhibit care not to excessively twist the cord attached to the load cell assembly. To
minimize cord twisting, allow the cord to hang vertically during installation.
4.
To reinstall the clamp, insert the clamp spacer between the base optics and the High-Pressure
Clamp. Be sure the spacer and clamp are firmly resting on their respective pins. Secure the High-
Pressure Clamp by using the long hex screws provided with the Digital Force Adapter.