background image

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 10

Warranty

Our equipment is warranted against defective material and workmanship (excluding batteries) for a 
period of three years from the date of shipment. Claims under warranty can be made by returning 
the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our 
option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our 
warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or 
use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any 
special, incidental or consequential damage.

SETTING UP VALVES

When setting up a valve it is important to correctly set the end stops. Use the PIECAL 434 to supply 
the 4 to 20 mA control signal to stroke the valve. Select “SOURCE” and the PIECAL 434 will use the 
internal power source for outputting current or switch to 2-WIRE SIMULATOR to stroke a valve using 
any pre-existing installed loop power supply as the power source.

Example:
1)  Disconnect the 4-20 mA control wires from the Current-to Pressure (I/P) converter or the actuator.
2)  Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 

2-Wire Transmitters

3)  Move the EZ-CHECK™ switch 

e

 to 

“4.00 mA”/”0.0%”

 and adjust the fully closed stop on the 

actuator.

4)  Turn the PIECAL 434’s knob 

slowly counterclockwise and verify that the actuator and valve 

don’t move. Repeat steps 3 & 4 until no movement is detected.

5)  Move the EZ-CHECK™ switch 

to 

DIAL

 and quickly back to 

“4.00 mA”/”0.0%”

 then turn the 

PIECAL 434’s knob 

clockwise. The actuator and valve should begin to move.

6)  Move the EZ-CHECK™ switch 

to 

“20.00 mA”/”100.0%” 

and adjust the fully open stop on the 

actuator.

7) Turn the PIECAL 434’s knob 

slowly clockwise and verify that the actuator and valve don’t 

move. Repeat steps 6 & 7 until no movement is detected.

8)  Move the EZ-CHECK™ switch 

to 

DIAL

 and quickly back to 

“20.00 mA”/”100.0%”

 then turn 

the PIECAL 434’s knob counterclockwise. The actuator and valve should begin to move

.

Stroking Valves

Additional Information 

This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration. 

Test Data is available for an additional charge.

Practical Instrument Electronics recommends a calibration interval of one year. Contact your local 

representative for recalibration and repair services.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for PIECAL 434

Page 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Page 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Page 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Page 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Page 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Page 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Page 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Page 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Page 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Page 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Page 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Page 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Reviews: