28
SETTNTC
SETTNTC
SETTNTC
SETTNTC
This command sets the NTC norm temperature to the given data. The bits 56 … 63 select the input
NTC channel, the bits 0…31 represents the norm temperature in a 32 bit signed integer format in
steps of 0.1°K.
The setpoint value must be within the borders defined by the minimum and maximum values given by
the
GETBNTCMIN
and
GETBNTCMAX
command.
The return value contains the actual resistance (See
GETBNTC
command).
GETRPTCMIN
GETRPTCMIN
GETRPTCMIN
GETRPTCMIN
The return value contains the minimum possible resistance value of the PTC channels. The lower 32bit
of the return value represents the data in a 32-bit integer format.
GETRPTCMAX
GETRPTCMAX
GETRPTCMAX
GETRPTCMAX
The return value contains the maximum possible resistance value of the PTC channels. The lower 32bit
of the return value represents the data in a 32-bit integer format.
GETRPTC
GETRPTC
GETRPTC
GETRPTC
This command returns the configured resistance of the given PTC input [0|1]. The lower 32bit of the
return value represents the data in a 32-bit integer format.
SETRPTC
SETRPTC
SETRPTC
SETRPTC
This command sets the PTC resistance to the given data. The bits 56 … 63 select the input NTC
channel, the bits 0…31 represents the resistance value in a 32 bit signed integer format.
The setpoint value must be within the borders defined by the minimum and maximum values given by
the
GETRPTCMIN
and
GETRPTCMAX
command.
The return value contains the actual resistance (See
GETRPTC
command).
GETIMAXMIN
GETIMAXMIN
GETIMAXMIN
GETIMAXMIN
The return value contains the minimum possible output current of the TEC. The lower 32bit of the
return value represents the data in a 32-bit integer format in steps of 0.01A.
GETIMAXMAX
GETIMAXMAX
GETIMAXMAX
GETIMAXMAX
The return value contains the maximum possible output current of the TEC. The lower 32bit of the
return value represents the data in a 32-bit integer format in steps of 0.01A.
GETIMAX
GETIMAX
GETIMAX
GETIMAX
This command returns the configured maximum output current of the TEC of the given channel [0|1].
The lower 32bit of the return value represents the data in a 32-bit integer format in steps of 0.01A.
SETI
SETI
SETI
SETIMAX
MAX
MAX
MAX
This command sets the PTC output current of the TEC to the given data. The bits 56 … 63 select the
TEC channel, the bits 0…31 represents the resistance value in a 32 bit signed integer format in steps
of 0.01A.
The setpoint value must be within the borders defined by the minimum and maximum values given by
the
GETIMAXMIN
and
GETIMAXMAX
command.
The return value contains the actual maximum current (See
GETIMAX
command).