14
Recommended Operating Conditions
Supply voltages
Min.
Typ.
Max.
Unit
HV+
0
-
+55
V
+5 V
+4.75
+5.0
+5.2
V
Trigger input
Min.
Typ.
Max.
Unit
EN
Differential
input low
threshold
-30
-100
mV
EP
Differential
input high
threshold
-30
0
mV
Pulse width
1
2
100
µs
Repetition rate
-
-
100
kHz
Channel selection
Min.
Typ.
Max.
Unit
CHSEL0 / CHSEL1 /
CHSEL2 / CHSEL3
input logic high
+2.4
+3.15
-
V
CHSEL0 / CHSEL1 /
CHSEL2 / CHSEL3
input logic low
-
2.1
+1.35
V
Address transition time
-
-
1.0
MHz
Absolute maximum Ratings (destroying limits)
Supply voltages
Min.
Max.
Unit
HV+
0
+55
V
+5 V
0
+5.2
V
Trigger input
Min.
Max.
Unit
Trigger signal voltage,
unterminated
0
+3.3
V
Trigger Signal:
Note that the maximum ratings for both
pulse width and repetition rate depend
on the actual high voltage supply (HV). See section “Power Losses” for guidance.
Summary of Contents for LDP-AV 16N45-40