Connecting the oscilloscope
The oscilloscope is supplied with the following cables and adaptors:
TA038 Current clamp adaptor
TA039 Screened twisted-pair cable
TA038 Current clamp adaptor
Use this adaptor to connect a current clamp to the oscilloscope. First fit the adaptor
to one of the BNC inputs on the oscilloscope's front panel, then insert the adaptor's
grounding plug into one of the ground sockets on the front panel.
It is important to ground the current clamp using the procedure just described. Most
current clamps have an internal metal case that readily picks up electromagnetic noise
from the environment, and if the case were not grounded, this noise would interfere
with the signal.
TA039 Screened twisted-pair cable
Use this cable to connect a signal directly to the oscilloscope. A positive signal will be
displayed when you connect the red BNC plug to the more positive signal and the
black BNC plug to the more negative signal. If you connect these the wrong way
round, the oscilloscope will not be damaged but the signal will appear inverted on the
scope display.
Insert the grounding plug into one of the ground sockets on the front of the
oscilloscope. This grounds the screen of the cable to prevent it from picking up
electromagnetic noise that might otherwise interfere with the signal.