6 Starting and Operating
24
Version: 1.1.0
PZ247E
P-63x Nanopositioner
NOTICE
Operating voltage too high or incorrectly connected!
Operating voltages that are too high or incorrectly connected can cause damage to the P-63x.
Operate the P-63x only with controllers/drivers and original accessories from PI.
Do
not
exceed the operating voltage range (p. 34) specified for the P-63x.
Operate the P-63x only when the operating voltage is properly connected; refer to "Pin
Assignment" (p. 36).
NOTICE
Uncontrolled oscillation!
Oscillation can cause irreparable damage to the P-63x. Oscillation is indicated by a humming
noise and can be caused by the following:
A change in the load and/or dynamics requires the servo control parameters to be
adjusted.
The P-63x is operated close to its resonant frequency, or with too high operating
frequency.
If you notice oscillation:
In closed-loop operation, switch off the servo mode immediately.
In open-loop operation, stop the P-63x immediately.
INFORMATION
The positive direction of motion is in the opposite direction of the cable exit.
INFORMATION
Sound and vibration (e.g., footfall, knocks) can be transmitted to the P-63x and can affect its
performance with regard to position stability.
Avoid sound and vibration while the P-63x is being operated.
6.2
Operating the P-63x
Requirements
You have read and understood the general notes on starting and operating.
You have read and understood the user manual for the electronics.
You have read and understood the user manual for the PC software.
You have correctly installed (p. 15) the P-63x, and connected it to the protective earth
conductor (p. 16).
The electronics and the required PC software were installed. All connections to the
electronics were made (refer to the user manual for the electronics).