5 Installation
P-616 NanoCube® XYZ Nanopositioner
PZ268D
Version: 1.1.0
17
Requirements
You have read and understood the general notes on installation (p. 15).
The P-616 is
not
connected to the electronics.
Tools and accessories
Suitable protective earth conductor: Cable cross section ≥0.75 mm
2
M4 protective earth screw set (p. 10) supplied for connecting the protective earth
conductor
Suitable screwdriver
Figure 4: Connecting the protective earth conductor (profile view)
1
Base body of the P-616
2
Flat washer
3
Lock washer
4
Screw
5
Cable lug
6
Protective earth conductor
Connecting the P-616 to the Protective Earth Conductor
1.
If necessary, attach a suitable cable lug to the protective earth conductor.
2.
Use the M4 screw (together with the flat and lock washers) to attach the cable lug of
the protective earth conductor to the threaded hole in the P-616 as shown in the profile
view.
3.
Tighten the M4 screw with a torque of 1.2 Nm to 1.5 Nm.
4.
Make sure that the contact resistance at all connection points relevant for connecting
the protective earth conductor is <0.1 Ω at 25 A.