5 Installation
M-126 Precision Microtranslation Stage
MP42E
Version: 3.0.0
19
5.3
Mounting the M-126 onto a Surface
Figure 4: Mounting plate on the bottom of the M-126
There is a mounting plate with two threaded holes (mounting holes) on the bottom of the M-
126. The M-126 is mounted on and underlying surface via this plate. Mounting is from above.
Access to the mounting holes is via two through-holes in the M-126's platform.
Requirements
You have read and understood the general notes on installation (p. 17).
You have provided a suitable underlying surface (see "Dimensions" (p. 36) for the
required position of the holes for the screws):
−
Two M4 threaded holes with sufficient depth are provided
Models with electric drive: Depth at least 18 mm
Model with manual drive: Depth at least 20 mm
−
For applications with large temperature fluctuations: The surface should have the
same thermal expansion properties as the M-126 (e.g., underlying surface made of
aluminum).
In addition, for models with electric drive:
If necessary, the motor housing and connectors of the M-126 were brought into a
suitable position (p. 18) for your application.
You have accounted for the space required to route cables without bending and
according to regulations.
Tools and Accessories
2 M4 socket head screws, included in the scope of delivery (p. 10)
Hex key, AF 3, included in the scope of delivery
Summary of Contents for M-126
Page 8: ......
Page 18: ......
Page 20: ......
Page 36: ......
Page 42: ...10 Technical Data 38 Version 3 0 0 MP42E M 126 Precision Microtranslation Stage M 126 DG1...
Page 43: ...10 Technical Data M 126 Precision Microtranslation Stage MP42E Version 3 0 0 39 M 126 PDx...
Page 44: ...10 Technical Data 40 Version 3 0 0 MP42E M 126 Precision Microtranslation Stage M 126 2S1...
Page 45: ...10 Technical Data M 126 Precision Microtranslation Stage MP42E Version 3 0 0 41 M 126 M0...
Page 52: ......