5 Installation
22
Version: 3.0.0
MP42E
M-126 Precision Microtranslation Stage
5.6
Connecting the Power Supply to the M-126
Connecting a power adapter is only necessary for the M-126.PDx models.
Requirements
The power cord is
not
connected to the power socket.
Tools and accessories
Supplied components:
−
24 V wide-range-input power supply
−
Adapter for the power supply connection; barrel connector 5.5 mm x 2.1 mm to
Mini XLR3 (f)
−
Power cord
If one of the components supplied for connection to the power source has to be
replaced: Use a sufficiently measured and certified replacement component. Details:
−
Power supply: Output 24 V DC, maximum output current 2 A
−
Power cord: Three wires, cable cross section at least 3 x 0.75 mm
2
(3 x AWG18),
maximum length 2 m
Connecting the power supply to the M-126
Connect the Mini XLR3 connector (f) of the adapter to the Mini XLR3 panel plug of the
M-126.
Connect the barrel connector of the adapter to the barrel connector socket of the
power supply.
Connect the power cord to the power supply.