6.6
Connecting the L-505.013212
NOTICE
Damage due to mechanical stress on the cable!
The cable will break from excessive or repeated bending. A broken cable will lead to failure
of the L-505.013212 or damage to the L-505.013212 or the electronics.
►
Install the L-505.013212 so that the cable has a bending radius of
≥
35 mm.
►
If necessary: Use cables that are approved for recurring mechanical loading (e.g., drag
chain cables).
Tools and Accessories
■
■
If necessary:
■
Drive/encoder cable to be used as extension cable and/or adapter cable, suitable for
the electronics
■
Four spacer bolts, part of the mounting hardware (included in the
■
If necessary: Suitable screwdriver for the locking screws of the connectors.
Requirements
✓
You have read and understood the
general safety instructions (p. 7)
✓
You have read and understood the user manual for the electronics used.
✓
You have installed the electronics properly.
✓
The electronics are switched off.
NOTICE
Damage due to incorrect connection of the L-505.013212!
Connecting unsuitable electronics or the wrong cable can lead to damage to the L-505.013212
or the electronics.
►
Make sure that the electronics support the drive type of the L-505.013212 and have been
configured accordingly.
►
Use cables from PI miCos only to connect the L-505.013212 to the electronics.
►
Pay attention to correct
Connecting the L-505.013212
1. If necessary: Remove the ESD protective caps from the connectors of the L-505.013212.
2. Optional, if you use an extension or adapter cable: Fix the space bolts onto the cable's
connectors.
3. Connect the L-505.013212 to the electronics:
■
Without extension/adapter cable: Connect the L-505.013212's drive connector to the
connector on the electronics.
■
With extension/adapter cable: Connect the cable to the L-505.013212's drive connector
and the drive connector on the electronics.
4. Secure the connectors against unintentional removal.
6 INSTALLATION
L505M0008EN
‒
6/17/2020
22
M O T I O N | P O S I T I O N I N G