background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

12620-09 and 12620-10 / 2715

 

Deleting files: 

Select the file with 



. Delete the file with the arrow key 

►. 

A  warning  asking  whether  you  really  want  to  delete  the  file 
will  be  displayed.  Press  the  confirmation  key 

  in  order  to 

confirm.  The  arrow  key 

◄  leads  you  back  to  the  "Files" 

menu. 

 

Opening a file and displaying it as a graph: 

Select the file with 



Confirm  your  selection  with 

.  The  file  will  be  opened  and 

displayed as a graph in the landscape format. 

 

 

The  graph  can  be  retraced  by  way  of  the  arrow  keys 

◄►. 

The green crosshairs on the graph indicate the current posi-
tion.  The  pair  of  values  of  the  graph  is  also  displayed  in 
green.  When  the  last  measuring  point  is  reached,  a  new 
segment  of  the  graph  will  be  displayed  automatically.  You 
can  exit  the  graph  with 

  and  return  to  the  file  selection 

window (Fig. 20

). ▲▼ can be used to change the measuring 

channel. 

 

5.4.5

 

Zero (3)

 

This  menu  item  enables  the  measurement  values  of  the 
measuring channels to be manually reset to zero. 

To  do  so,  the switches  of  the  desired  channels  are  set  from 

 off  to 

 on  by  way  of  the  arrow  keys 

◄► 

(Fig. 20).  

You  can  exit  the  submenu  and  return  to  the  menu  by  way 
of 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.4.6

 

Calibration (4)

 

In the case of the one-point calibration (4.1), the value that is 
currently  measured  via  a  channel  (actual)  is  shifted  to  an-
other value (set) by way of a constant (Fig. 23). 

In the case of the two-point calibration (4.2), each of the two 
values that are measured is assigned to a new value. 

Before  the  new  calibration  data  are  stored  in  the  sensor,  an 
"attention"  message  is  displayed.  Confirming  the  message 
with 

 stores the values, whereas 

◄ cancels the storage.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 21: Representation of a file in the form of a graph 

Fig. 22: Display of the submenu "Zero" for the Cobra4 Mobile-

Link 2 

Fig. 23: Display

 

of the submenu "Calibration" for the Cobra4 

Mobile-Link 2 

Zero 
 
Menu

 

ax

 

0.00 g

    

  

 

 

 

ay

 

0.34 g  

  

 

 

 

az

 

0.00 g

    

  

 

 

 

 

 

 

 

 

Menu 

 

One Point 
 
Menu

 

Channel    

 

    G

 

 

   

 

 

Ref.              -0.1 

 

μS/cm 

 

Actual             0.3     μS/cm 

 

 

 

 

 

Menu > Calibration

 

 

 

 

 

-1 

 

 
-2 
 

ax / g 

t / s 

1       2       4       5      6       7       8       9      10    11 

Acceleration

 

0.95 g 
0.8 s 

Summary of Contents for 12620-09

Page 1: ...nsor Setup 1 2 5 4 3 Datalogger Setup 2 1 5 4 4 Datalogger Files 2 2 5 4 5 Zero 3 5 4 6 Calibration 4 5 4 7 Setup Date Time 5 1 5 4 8 Setup Battery status 5 2 5 4 10 Setup Information 5 4 5 4 11 Setup...

Page 2: ...supply unit 3 FUNCTIONAL AND OPERATING ELEMENTS 3 1 Operating elements on the top The Cobra4 Mobile Link 2 is equipped with various elements for the measurement data display operation and status indi...

Page 3: ...nity of the unit After a complete shut down of the unit it must be switched off and on again by way of the ON OFF switch operating ele ment If the unit does not react keep the ON OFF switch pressed fo...

Page 4: ...performed up to three times 5 3 Measurement data recording process A measurement can be started and stopped by pressing the confirmation key The measurement data recording proc ess is indicated by a...

Page 5: ...The file name consists of the date and a running counter YYMMDDxx mli xx 00 99 Ensure that the date and time are set in the Cobra4 Mobile Link This can be checked and if needed changed in the Setup m...

Page 6: ...Unit Conductivity 12632 00 It enables the automatic temperature compensation of a pH or conductivity measuring channel e g T 1 with regard to a temperature input or the manual input of a known temper...

Page 7: ...e following Cobra4 Sen sor Units Cobra4 Sensor Unit Radioactivity 12665 00 It is used to change the counter tube voltage in order to measure the counter tube characteristics Gate time This setting opt...

Page 8: ...imer Counter 12651 00 5 4 3 Datalogger Setup 2 1 The following changes are possible Selection of the single value measurement mode by way of a switch This ensures that only one value will be re corded...

Page 9: ...u item enables the measurement values of the measuring channels to be manually reset to zero To do so the switches of the desired channels are set from off to on by way of the arrow keys Fig 20 You ca...

Page 10: ...played in the header The automatic rotation of the device display can be switched on and off Use the confirmation key in order to confirm a new setting 5 4 10 Setup Information 5 4 Here you can find g...

Page 11: ...es minutes and seconds This menu item can only be used in combination with a Mo bile Link 2 GPS 12620 20 and GPS activated See also chapter 5 4 9 Coordinates will not be displayed unless satellites ha...

Page 12: ...f the battery decreases strongly it is possible to change the battery Please note that damage resulting from a battery change are not covered by the warranty You can send the Mobile Link to the repair...

Page 13: ...Unit Energy 12656 00 Cobra4 Sensor Unit Forceplate 12661 00 Cobra4 Sensor Unit Radioactivity 12665 00 Cobra4 Sensor Unit Sound Level 12669 00 Cobra4 Sensor Unit Weather air pressure humidity temperatu...

Reviews: