Service Manual PhysioFlow Lab1 

  9 / 20

 

Version of Monday, 10 July 2017 

 

Symbols and Marks 

Symbol 

Signification 

 

In the manual

: Important information to consider for an efficient and 

optimized use of the system.  
 

On the device panel

: Led indicator to mention that an error has been 

detected.  If  the  LED  is  on  please  contact  PhysioFlow®  support 
platform (

[email protected]

 

 

In  the  manual  this  symbol  indicates  that  one  or  several  conditions 
could  damage  the  equipment  itself  and/or  impact  patient’s  and/or 
user’s and/or environment’s safety.   

 

PhysioFlow®  power  light:  When  the  LED  is  on  then  the  device  is 
correctly powered.  

 

Connections linked to the patient are BF Type. 

 

« CE »  mark  followed  by  the  notified  organization’s  registration 
number. It ensures that the device meets the essential requirements 
of the European directive 93/42/CEE on medical devices. 

 
 

 

Summary of Contents for Lab1

Page 1: ...730 Folschviller France T l 33 0 3 72 82 50 00 Fax 33 0 1 30 74 46 48 support physioflow com Service Manual PhysioFlow Lab1 1 20 Version of Monday 10 July 2017 PHYSIOFLOW Lab1TM Service Manual Monday...

Page 2: ...e 4 Measuring Technique Information 6 Warnings 7 Precautions 8 Symbols and Marks 9 2 Overall presentation 10 3 Interfaces 11 Description of the electronic unit 11 Description of cables 12 Description...

Page 3: ...ng this manual is required to ensure a proper use of the medical device Basic technicians and maintenance staff education associated to reading this manual is required for any intervention on the medi...

Page 4: ...or the PhysioFlow application it equates to a 1 5m distance around the patient himself or of the surface he is in contact with in any direction Conditions of Use The PhysioFlow Lab1 is a noninvasive h...

Page 5: ...ulmonary causes of acute dyspnea Optimization of atrioventricular interval for patients with A V sequential cardiac pacemakers Patients with need of determination for intravenous inotropic therapy Pos...

Page 6: ...differences in thoracic electrical impedance are essentially created by changes of velocity and blood volume in the aorta Unlike other trans thoracic bio impedance systems PhysioFlow System does not r...

Page 7: ...r in the presence of flammable anesthetics Sensors must be only positioned on the skin and nowhere else Disposal of this product and or its accessories shall be made in accordance with all locally enf...

Page 8: ...e electromagnetic emissions and reduce the immunity of the device PhysioFlow devices must not be sterilized FOR ACCURATE MEASUREMENTS IT IS ESSENTIAL THAT THE OPERATOR UNDERSTANDS THE DIFFERENCE BETWE...

Page 9: ...on that an error has been detected If the LED is on please contact PhysioFlow support platform support physioflow com In the manual this symbol indicates that one or several conditions could damage th...

Page 10: ...cted to a PC compatible computer through a RS 232 C serial link It is based on the principles of impedance cardiography The electronic unit performs impedance signal generation reception of chest impe...

Page 11: ...de Petit Ebersviller 57730 Folschviller France T l 33 0 3 72 82 50 00 Fax 33 0 1 30 74 46 48 support physioflow com Service Manual PhysioFlow Lab1 11 20 Version of Monday 10 July 2017 3 Interfaces Des...

Page 12: ...rvice Manual PhysioFlow Lab1 12 20 Version of Monday 10 July 2017 Description of cables A Electrical RS 232 serial link cable to be used with a laptop computer only Type zero modem with SUB D9 Male Fe...

Page 13: ...u stade Petit Ebersviller 57730 Folschviller France T l 33 0 3 72 82 50 00 Fax 33 0 1 30 74 46 48 support physioflow com Service Manual PhysioFlow Lab1 13 20 Version of Monday 10 July 2017 Description...

Page 14: ...physioflow com or your local dealer No ECG and no impedance signal are displayed verify that the patient cable is connected on PhysioFlow and that the serial connection is OK Replace electric RS 232...

Page 15: ...completed on the computer The external data acquisition system has to be calibrated adjustment of scales to reproduce measurements displayed on the PhysioFlow Computer screen To perform this task sta...

Page 16: ...flow com With special authorisation of the manufacturer or the dealer the unit s embedded processor software can be upgraded To perform this operation it is necessary to replace the EPROM First stage...

Page 17: ...ly 2017 Storage and cleaning Out of operation periods the PhysioFlow unit must be stored at ambient temperature between 40 C and 70 C humidity included between 10 and 100 and an atmospheric pressure i...

Page 18: ...ndix A Legislation Computer must comply with regulations concerning medical devices Desktop computer or laptop with leakage current under 0 1 mA Please refer to the manufacturer information Please con...

Page 19: ...ctors Back panel Female SUB D9 serial link to PC Switcher filter fuses module Schaffner Front panel Specific patient connector 6 leads ground Gilded contacts Measurement current Intensity of 3 6 mA pe...

Page 20: ...ation 15 seconds averaging patient at rest Parameter Unit Low value High Value Variability Heart rate HR bpm 30 215 1 Stroke Volume SV mL 0 220 5 Stroke Volume Index SVi mL m 0 100 5 Cardiac Output CO...